[:Sense]:Bandwidth Subsection - Agilent Technologies N9340B Programmer's Manual

Spectrum analyzer
Table of Contents

Advertisement

5
Command Reference

[:SENSe]:BANDwidth Subsection

Default Unit:
82
Type of Averaging for Measurements
[:SENSe]:AVERage:TYPE LOGPower|POWer|VOLTage
[:SENSe]:AVERage:TYPE?
Use this command to select the averaging type for the
measurement:
LOG POWER - Averages the data as appropriate for the
logarithmic scaled y- axis.
Power - Converts the scale from logarithmic to linear power
units, and then averages the power trace data.
Voltage - Averages the data as appropriate for the linearly
scaled y- axis scale.
*RST:
log power
Key access:
BW/SWP > Average Type
Resolution Bandwidth
[:SENSe]:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:BANDwidth|BWIDth[:RESolution]?
Specifies the resolution bandwidth.
Example:
BAND 1 kHz
Range:
10 Hz to 3 MHz
Hz
Key access:
BW/SWP > RBW Auto Man
N9340B Programmer's Guide