ESA Functional Tests
Reference Level Accuracy
Reference Level Accuracy
Test Limits
Reference Level
dBm
dBmV
–30
21.76
–20
31.76
–10
41.76
–40
11.76
–50
1.76
–60
–8.24
–70
–18.24
Test Description
A 50 MHz CW signal is applied to the Input of the analyzer. The amplitude of the source
and the analyzer's reference level are decreased in 10 dB steps. The analyzer marker
functions are used to measure the amplitude difference between steps. Reference Level
Accuracy is tested in both Log and Linear Scale Modes. Most of the error is contributed
from the output attenuator inaccuracy in the synthesized sweeper and not the analyzer.
Item
Signal Sources
Synthesized Sweeper
Output Level Accuracy
0 to –15 dBm: ±1.0 dB
–16 dBm to –63 dBm: ±1.4 dB
≤ –64 dBm: ≥2.0 dB
Adapters
Type-N (m), to
BNC f)
Type-N (m), to
APC 3.5 (f)
Cables
Type-N, 152-cm
(60-in)
BNC, 122-cm (48-in)
172
Minimum
Maximum
(dB)
(dB)
Reference
Reference
–1.40
1.40
–1.40
1.40
–1.40
1.40
–1.40
1.40
–1.40
1.40
–2.0
2.0
Critical Specifications
(for this test)
Recommended
Agilent Model
8340A/B or 836XX
Series
1250-1476
1250-1745
11500D
10503A
Chapter 4