Power Interrupted; Pumps Off Over 3 Minutes - Gambro Cobe Spectra Essentials Manual

Table of Contents

Advertisement

POWER INTERRUPTED!

POWER INTERRUPTED! Continue with
previous procedure (YES/NO)?
Alarm LED is red. Pumps stop. Return line valve closes.
Probable Cause
Power restored after a power interruption

PUMPS OFF OVER 3 MINUTES.

PUMPS OFF OVER 3 MINUTES.
Start saline drip.
Alarm LED is Red. During a platelet procedure, centrifuge speed
decreases to 1800 rpm.
Probable Cause
Pumps paused for more than 3 minutes
Return line valve manually closed during a Single-Needle
Return phase
®
COBE
Spectra™ Apheresis System • Essentials Guide
Respond to question: "Continue previous procedure (YES/NO)?" by pressing either YES or NO.
If you press YES, restart the interrupted procedure by pressing CONTINUE.
If you press NO, follow the instructions on the display.
CONTINUE
If the pumps are to remain paused, start a saline drip. When ready to resume the procedure, restart the
pumps by pressing CONTINUE.
Manually open the return line valve. Restart the pumps by pressing CONTINUE.
Action
To override the alarm, press CONTINUE.
This alarm occurs only during Run mode.
Warning: DO NOT unload the disposable tubing set if the donor
or patient is connected to the Spectra system, or the donor or
patient may receive anticoagulant and other unwanted fluids
through the access and return needles.
Action
General Alarms and Troubleshooting
12-49