RETURN PRESSURE HIGH! Check return line and needle.

RETURN PRESSURE HIGH!
Check return line and needle.
Alarm LED is red. Pumps stop.
Probable Cause
Blockage in the return line or needle
Return pressure alarm limit too low for current operating
conditions
Buildup of pressure in the system caused by an alarm
condition, or operator action caused the centrifuge to stop
Occluded return vein
®
COBE
Spectra™ Apheresis System • Essentials Guide
CONTINUE
Check for kinks in or clamps on the return line, or clots in the return needle. Ensure the correct clamp was
closed before Rinseback mode.
During Single-Needle ELP, LRS, or LRST Prime mode, ensure the clamp between access manifold and the
Single-Needle "Y" Manifold is open.
Increase the return alarm limit if you changed it earlier in the procedure. Ensure the return needle is not
infiltrated.
Release the pressure in the system by performing one of the two sets of steps below:
Option 1. Release the pressure into the disposable tubing set as follows:
1
Close the pinch clamp on the return line. Open the roller clamp on the return saline line.
2
Manually pull out the return line valve. The pressure will vent to the saline bag.
3
Close the return saline line roller clamp and open the pinch clamp on the return line.
4
Press CONTINUE.
– or –
Option 2. Release the pressure into the waste bag as follows:
1
Open the waste divert valve to relieve pressure: Press VALVE, press 3 to choose the waste valve, then
press 2 to select the return divert position.
2
Close the waste divert valve to return to the operating position, then press CONTINUE.
Check the return needle site.
This alarm occurs when pressure in the return pressure sensor measures
above specified limit.
Action
General Alarms and Troubleshooting
12-55