Software Options and Configuration
Collect/Replace Valve
Press VALVE, then press 4 to select "Collect." The following screen
appears:
Collect Valve (now ________):
1=Collect, 2=Return, 3=Load.
The screen displays "now Collecting," "now Returning," or "now Open,"
depending on the current position of the collect/replace valve.
To open the collect/replace line to the collection bag, press 1.
To return the contents of the collect/replace line to the
donor/patient, press 2.
To open the collect valve so you can load or remove the disposable
tubing set, press 3.
During TPE and RBCX procedures, when the collect/replace valve is in
the "Collect" position, the system pumps replacement fluid through the
line to the patient.
During collection procedures, when the collect/replace valve is in the
"Collect" position, the system pumps the contents of the collect line to
the collection bag.
Plasma/RBC Valve
Press VALVE, then press 5 to select "Plasma." The following screen
appears:
Plasma Valve (now ________):
1=Collect, 2=Return, 3=Load.
The screen displays "now Collecting," "now Returning," or "now Open,"
depending on the current position of the plasma/RBC valve.
5-16
To open the plasma line to the collection or plasma/RBC waste bag,
press 1.
To return plasma to the donor/patient, press 2.
RBCX procedures: To return RBCs to the donor/patient, press 2.
To open the plasma/RBC valve so you can load or remove the
disposable tubing set, press 3.
During TPE procedures, when the plasma/RBC valve is in the collect
position, the system pumps plasma to the plasma waste bag.
During RBCX procedures, when the plasma/RBC valve is in the collect
position, the system pumps RBCs to the RBC waste bag.
During collection procedures, the plasma/RBC valve is in the collect
position only while plasma is collected and pumped to the plasma bag.
All Valves
Press VALVE, then press 6 to select "All valves." The following screen
appears:
1=Close All Valves, 2=Open All Valves.
To close all valves, press 1. The return, RBC, waste or return waste,
plasma bag, and collection bag lines close.
To open all valves so you can load or remove the disposable tubing
set, press 2.
This option is not available during Run mode.
®