This service manual provides the information writing by the manufacturer and carried out needed to install the AK 95 or AK 95 S by appropriately trained and suitable monitor and how to carry out maintenance qualified people;...
Publication no. Denomination HCEN9095, Rev. 10.2006 AK 95 / AK 95 S - Service Manual The complete AK 95 / AK 95 S - Service Manual, HCEN9095, consists of the following separate chapters: HCEN9133 Rev. 01. 2004 Installation Guide HCEN9134 Rev. 10. 2005 Technical Description HCEN9369 Rev.
This installation Guide provides the information needed to install both AK 95 and AK 95 S. The text from this point forward only states AK 95 S. AK 95 / AK 95 S Service Manual HCEN 9133 Rev. 01. 2004...
Physical Data When unpacking, the equipment should be The AK 95 S can be installed in most environ- checked. If the equipment is in any way dam- ments if certain conditions as the temperature, aged, proper operation cannot be guaranteed.
Installation Guide Mains Voltage and Power Consumption Tools The AK 95 S must be plugged into a grounded The tools listed below are needed during the power point not more than 3 metres distant. installation procedure. Mains voltage: • Tube cutter, 113 500 084 100 V(at 50 Hz or 60 Hz) •...
1,5 mA. The AK 95 S charge the battery as long as the AC/DC con- Main switch, verter is switched on, i.e. AK 95 S itself does not ON/OFF need to be switched on. The battery is tested during every functional...
Move the cover upwards and then back- wards. Pull the cover to the right just 10 mm (if you are positioned in front of the AK 95 S). Turn the cover backwards counter-clockwise to pass the circuit boards and the cover can be removed.
Installation Guide Installation of RP 95 S NOTE! This instruction is for AK 95 S equipped with RP 95 S at delivery. RP 95 S can also be ordered and installed as a kit, K23054. Refer to rebuilding instruction K23055.
Page 14
2. Fasten the cable even on the back part of machine with the enclosed cable holder - use an existing screw. AK 95 / AK 95 S Service Manual HCEN 9133 Rev. 01. 2004...
Fuse: T15 A Running time: 30 minutes The charge indicator on the power supply is lit when the AK 95 S is equipped with the back-up battery. The machine must be connected to mains voltage and the mains switch must be on.
• That the drain tube is properly connected to the machine and placed with an air gap between If a power failure occurs when AK 95 S is work- the drain tube and the drain. ing, the same segment in the TIME display is flashing.
Page 17
Installation Guide This page is intentionally left blank 1:10 AK 95 / AK 95 S Service Manual HCEN 9133 Rev. 01. 2004...
Gambro Service representative. This record is to be signed and filed by the Service Engineer responsible for the installation. Date......Signature................ Serial number: ........1:11 AK 95 / AK 95 S Service Manual HCEN 9133 Rev. 01. 2004...
Page 19
Installation Guide This page is intentionally left blank 1:12 AK 95 / AK 95 S Service Manual HCEN 9133 Rev. 01. 2004...
Introduction General AK 95 S operates in either acetate- or The AK 95 S is designed to be used as a single bicarbonate mode. In bicarbonate mode the patient machine to perform conventional hemo- monitor mixes dialysis fluid from water and two dialysis.
For information about the use and functions of the parameters in level 4, see Ace tate AK 95 Operators Manual, HCEN 9542 or AK 95 S Operators Manual, HCEN 9794. Figure 1. AK 95 S. AK 95 / AK 95 S Service Manual HCEN 9134 Rev. 10.2005...
+130 mmHg. Normaly during treatment INVA is open. The main function of the temperature transduc- ers in the AK 95 S is to measure the dialysis fluid temperature and control the heat regula- tion. The regulating temperature transducer, , is located after the heater.
Page 29
Blood leak detector Expansion chamber Degassing chamber Deairating chamber Variable flow unit (Stepper motor) Line clamp unit Clamp house Dialyzer connection Figure 3. AK 95 S - Fluid part AK 95 / AK 95 S Service Manual HCEN 9134 Rev. 10.2005...
Page 30
The acid/ MM 95 B, Ac Cart Mixing Chamber Mixing Chamber Feeding Pumps Cond. Cond.Cell "B" Cell "A" FLVA pH-SENSOR Figure 4. MM 95 - Flow diagram AK 95 / AK 95 S Service Manual HCEN 9134 Rev. 10.2005...
Page 31
Technical Description Feeding Pump Mixing Chambers B Heat Exchanger FLVA Conductivity Cell Figure 5. MM 95 2:10 AK 95 / AK 95 S Service Manual HCEN 9134 Rev. 10.2005...
HPG transducer is above or equal to +550 mmHg calculation since it varies with temperature, (presetable). approximately +1,8 % per °C. 2:11 AK 95 / AK 95 S Service Manual HCEN 9134 Rev. 10.2005...
Page 33
During calibration of the UF measuring unit, the taration valve is closed. TAVA is also closed when the protective system demands a minimum of UF. 2:12 AK 95 / AK 95 S Service Manual HCEN 9134 Rev. 10.2005...
230 V 110 V 100 V will be an attention. With UFD installed When the filling phase is finished the AK 95 S Heat will give an attention that requests you to place Heat Citric acid 20% 52 the B-tube back in the monitor. When this is...
(approx.. +130 mmHg). which detects when the tubes are disconnected from the safety coupling. The AK 95 S is also equipped with a pair of Conductivity measuring unit mechanical switches that sense if the dia connec- The unit includes the following components: tors are attached to the bypass block.
Technical Description CPU board Blood leak detector AK 95 S is equipped with one Control CPU board The blood leak detector includes the following and one Protective CPU board. components: The CPU boards comprises among other things • Housing with cover.
Technical Description Rear printed circuit boards Figure 8 . AK 95 S CPU-boards and Driver boards. 2:16 AK 95 / AK 95 S Service Manual HCEN 9134 Rev. 10.2005...
Dialyzer Patient Venous blood line Venous drip chamber Priming Air detector detector Venous line clamp Venous pressure transducer Figure 9. AK 95 S - Blood Flow Diagram. 2:17 AK 95 / AK 95 S Service Manual HCEN 9134 Rev. 10.2005...
If the event of an alarm, the blood flow is stopped. If the venous pressure is too high, the cause may • An obstruction after the venous drip chamber. • A change in the patients position. 2:18 AK 95 / AK 95 S Service Manual HCEN 9134 Rev. 10.2005...
• Housing with covers (mounted on the front Arterial pressure transducer plate). AK 95 S can be equipped with a second pressure • 2 solenoids (as one unit). transducer, named Arterial Pressure Trans- •...
1 numerical display (LED). • 2 knobs. The panel interfaces to the protective driver- board, via 3 board connectors and 3 flat cables. Figure 12. Operators panel 2:22 AK 95 / AK 95 S Service Manual HCEN 9134 Rev. 10.2005...
The module includes con- the device to determine if it works or not. trol and protective systems. The device has to be sent to Gambro for repair The control system measures the pressure and incase of malfunction. pulse wave to calculate the systolic, diastolic, mean blood pressure and pulse rate.
Power supply to heater relay board Control CPU board External Fuses: 12 AH T12AH 250AC HALT BACH BACO Contact Fan Intake Mains Switch Figure 13. AK 95 S, AC/DC-unit 2:24 AK 95 / AK 95 S Service Manual HCEN 9134 Rev. 10.2005...
DC supplies to the the heating element. monitor. In a mains power failure situation, the AK 95 S •Two +5 volt supplies, one for the control system gives an intermittent buzzer alarm and performs and the other for protective digital electronics.
Page 47
Technical Description This page is left intentionally blank. 2:26 AK 95 / AK 95 S Service Manual HCEN 9134 Rev. 10.2005...
Standards ..........3:16 H CEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations Program version 12.xx...
Page 49
Heparinization stops before end of treatment 0.00 to 9.59 h Counter pressure: Maximum 400 mmHg Accumulated heparin 0 to 999.9 ml volume: HCEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations Program version 12.xx...
Page 50
High alarm limit, can be 40 - 200 bpm (150 bpm) preset (Default) Meets ANSI/AAMI SP-10 (1992). Mean error ±5 mmHg. Standard deviation 8 mmHg H CEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations Program version 12.xx...
Page 51
- , 20 to 40 mmol/l, (±6 mmol/l) Measuring range: 13 to 16 mS/cm Alarm limits: ±5% of the calculated conductivity set value HCEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations Program version 12.xx...
Page 52
Maximum 20 min, not intended for when not in use overnight disinfection! Total time 30 minutes (with UFD-kit 48 minutes) including 10 min. contact time H CEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations Program version 12.xx...
Page 53
Heat Citric acid 20 % H eat C leanCart Without UFD Heat Heat Citric acid 20 % H eat C leanCart HCEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations Program version 12.xx...
Page 54
It is possible to use water with higher conductivity if it consists mainly of sodium salts. This may however affect the accuracy of the fluid composition. H CEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations Program version 12.xx...
Page 55
10 µA (100 V) 10 µA (110 V) 10 µA current: (230 V) Note: All leakage currents specified are without external equipment connected to the AK 95 S. HCEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations Program version 12.xx...
Page 56
Low level max output -5.0 V voltage: Max output current: ±5 mA RS-422/Current loop: Typ current: 20 mA Max current: 50 mA H CEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations Program version 12.xx...
Page 57
Battery back-up Battery back-up of power 24 volt, 6,5 Ah supply Running time >30 minutes Fuse T 15 A HCEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations 3:10 Program version 12.xx...
Page 58
Supervision of the stop time of the blood pump Alarm: The attention “BLOOD PUMP STOP TIME EXPIRED, Start Blood pump”. H CEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations 3:11 Program version 12.xx...
Page 59
Measuring range: 1 - 9.9 pH, measured in the dialysis fluid before dialyzer. Accuracy: ±0.2 pH Alarm limits: 5.0 to 9.0 pH HCEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and specificat ions 3:12 Program version 12.xx...
Page 60
Approx. 1270 mm (without infusion stand). Weight: Approx. 62-78 kg (depending on options). Infusion stand Maximum total load: 2 kg References Assembly Drawing K23000 H CEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations 3:13 Program version 12.xx...
Page 61
Metals Stainless steel SS2343 Stainless steel SS2353 Stainless steel SS2562 Titanium Platinum Others Carbon Ceramic, Steatite 221 Ceramic, Aluminium oxide Glass HCEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations 3:14 Program version 12.xx...
Page 62
Ambient Temperature range -20 to +70 °C Relative Humidity range 10 to 96% RH Air Pressure range ( atm. 500 to 1060 hPa Pressure ) H CEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations 3:15 Program version 12.xx...
Page 63
EN 1060-1 Non-invasive sphygmomanometers Part 1: General requirements EN 1060-3 Non-invasive sphygmomanometers Part 3: Supplementary requirements for electromechanical blood pressure measuring systems HCEN9369 Revision.10.2005 AK 95 S Service m anual - Technical data and spec ifi cations 3:16 Program version 12.xx...
Page 64
Technical Data Chapter 3 Technical data and specifications for AK 95 AK 95 HCEN9135 Rev. 10.2005...
Page 65
Technical Data Machine specification AK 95, K18022 Blood flow control Double needle Flow rate: 0 and 20 - 500 ml/min (-150 mmHg prepump pressure, pump segment diameter 7.9 mm) Flow accuracy: ±10 ml/min or ±10 %, whichever is largest Accumulated blood volume: 0 - 999 litres Volume accuracy: ±10 %...
Page 66
Technical Data Air detection Detection method: Ultrasonic detector placed at the venous drip chamber. The detector has a two channel structure and the function of the detector is tested at the function test made by the microcomputers. Drip chamber size: Diameter 22 mm Sensitivity: Bubbles larger than 1 µl will be trapped by the drip chamber.
Page 67
Technical Data Safety guard switch (SAGS): Increasing: -59 mmHg, ±7 mmHg Decreasing: -74 mmHg, ±7 mmHg Profiling UF-rate 0.0 to 4 l/h , Acetate mode 130 to 160 mmol/l , Bicarbonate mode 130 to 160 mmol/l , Bicarbonate mode 20 to 40 mmol/l Dialysis fluid preparation Pressure regulators: After pressure regulator...
Page 68
Technical Data Bicarbonate: , 130 to 160 mmol/l, (±6 mmol/l) - , 20 to 40 mmol/l, (±6 mmol/l) Measuring range: 12 to 16 mS/cm Alarm limits: To 5 % of the calculated conductivity set value pH-meter: 1 – 9.9 pH, measured in the dialysis fluid before dialyzer. ±0.2 pH Accuracy: Alarm limits:...
Page 69
Technical Data Hypochlorite Program (presetable) Concentration of disinfectant 10 % available chlorine Concentration in machine 0.5 %; i.e. diluted 1 + 19 (1:20) Volume Approx. 145 ml (with UFD-kit approx. 153 ml) Contact time between treatments 10 minutes Contact time overnight or when not in Maximum 20 min, not intended for overnight disinfection! Total time...
Page 70
Technical Data Auto Heat disinfection: One of two alternatives for auto heat disinfection can be selected. Total Time [min] AK 95 Disinfection 110 V 100 V 230 V programs Heat With UFD installed Heat CleanCart Heat Without UFD Heat CleanCart Temperature: +93°C (measured after heating rod) ≥80°C (measured in the outlet before the heat exchanger )
Page 71
Technical Data Mains plug: For 100/110V, Hospital grade, earth plug, type IEC 83 A5- For 230V, Dual earth system plug, type IEC 83 C4 100 µA (100V) Maximum earth leakage 150 µA (110V) current: 205 µA (230V, 50 Hz ) 400 µA (230V, 60 Hz ) 9 µA (100V) Maximum patient leakage...
Page 72
Technical Data Dimensions and weight Width: Machine 350 mm, stand 625 mm. Depth: Machine 420 mm, stand 680 mm. Height: 1270 mm. Weight: 62 kg Infusion stand Maximum total load: 2 kg Materials in contact with dialysate, concentrates and water Polymers Silicon rubber Santroprene...
Page 73
Technical Data Environmental data – transportation and storage -20 to +70 °C. Ambient temperature: Relative humidity: 10 – 96 % RH Air pressure: 500 to 1060 hPa If transportation or storage time is more than 15 weeks, the environmental data relating to the operation has to be followed.
Page 74
UF diagnostics ........4:24 Valves diagnostics ........ 4:24 Logging mode ..........4:25 Logging mode flow diagram ....4:26 Error buffers in Logging mode .... 4:27 Arrays ........... 4:28 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 75
Time (Remaining treatment time). - Heparin, Priming, Auto Priming (Priming functions) and Rinseback. - Air detect, Attention, Technical alarm, pH, and Blood leak. - Profiling and BPM. AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 76
4. Stand-by group active mode is indicated by a lit button. For These buttons can be used for direct control and draining of the AK 95 S, press the Rinse/Drain stand-by of certain functions. button twice. Is used to stop and start the blood pump.
Page 77
Treatment Treatment, fluid bypass, Isolated UF Green Yellow Green Green Green Green Standby, fluid OK/NOT OK Treatment, fluid NOT OK Yellow Yellow Green/Orange Green/Orange Orange Orange Bypass AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 78
Blood flow, Venous pressure, TMP, and PROM-configurations. This will prevent the UF volume, when “Select” is pressed: possibility of mixing PROMS from different program versions when installing/updating the AK 95 S. If the configuration is invalid, an error Bl:ml/min V :mmH TMP:mmH UFV:l 0.00 code will be issued.
Page 79
Service Technicians Guide Technical error handling For more information about the error message refer to the Error Code List for AK 95 S, HCEN 9091. Note! It is important that the program version on the Error Code List corresponds with the AK 95 S PROM.
Page 80
When desired mode is blinking, confirm with the button. START • To store values or modes, press STOP • To step backwards and finally exit from the menu, press AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 81
Service Technicians Guide Preset mode In the Preset mode the set up of AK 95 S is confirmed. Following choices are set in preset mode: • Start up values • Configuration • Set values and limits • Override times • Auto priming parameters •...
Page 82
Select Ext with the SET knob and press SELECT key. SELECT key and following display shows up: The AK 95 S Presets are based on the concept of groups. The main groups are: SELECT EXTERNAL PRESET TYPE • 0 Start up...
Page 83
When ready the machine returns to the first display in Manual preset mode. Note! When selecting Default preset, AK 95 will perform a Chem Rinse at the next start up. 4:10 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 84
The time when the venous clamp is open. In order to expell any air bubbles in the venous line. 13 Auto priming FM active Avail / Not Avail 4:11 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 85
Time build time drop time Venous Level line fill adjust- ment Flush 2 Pressure test Flush 1 Priming Fill Preset Time parameter 3,4,5 6,7,8 6,10 6,11 6,12 4:12 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 86
MAINTENANCE SELECTION Preset Service Example Following example will show how to preset the Select Preset and following text is displayed: AK 95 S for Gambro BiCart together with A- concentrate 201. INITIALIZING PRESET MODE Bicarbonate concentrates SELECT PRESET TYPE Manual Ext Default 1.
Page 87
COONa NaCl NaCl MgCl MgCl CaCl O with MgCl CaCl O with CaCl NaHCO NaHCO NaCl NaCl COONa COONa O with CH COONa BiCart BiCart BiCart BiCart 4:14 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 88
MAINTENANCE SUPPORT SELECTION Logging Service/Preset Select Service/Preset, set the service code and HOLD key, to step backward. enter the Service mode. Following displayshows up: MAINTENANCE SELECTION Preset Service 4:15 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 89
(pumps stopped) in External mode. •In Internal AK 95 S maintains a flow. SELECT AD-CONVERTER CPU C CPU P CPU W MM 95 See AD system calibration. 4:16 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 90
CONDUCTIVITY A TRANSDUCER CONDUCTIVITY P TRANSDUCER CONDUCTIVITY B TRANSDUCER To step through the Pressing the Select knob Selection diagram (a loop), opens the calibration display. use the SET knob. 4:17 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 91
Stand-alone Erasing Chemical disinfection indication REMOVE DISINF/DECALC ATTENTIONS KEEP RINSE/DRN pressed and wait Chemical disinfection indication is OFF! Removing DISINF/DECALC ATTENTIONS Disinfection and Decalcification attentions are removed 4:18 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 92
Press SELECT to confirm. I2C EEPROM SCAN COMPLETED DIAGNOSTIC ERROR BUFFER ERASED Name of the transducer: XXXXX EEPROM ADDRESS ERROR Update Escape XXXXX EEPROM NOT INITIALIZED Init Escape 4:19 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 93
Press SELECT to view lamps PUMP DIAGNOSTIC MENU Any lamp not lit is defect HEPARIN PUMP DIAGNOSTIC ART FREQ Counters 00000 00000 COUNTERS SETFREQ EST-ACC ACC 00000 00000 00000 4:20 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 94
Service Technicians Guide FM diagnostics DIAGNOSTICS FM MENU Var flow Pumps UF Valves See “Variable flow-diagnostics” “Valves-diagnostics” “UF-diagnostics” PUMP DIAGNOSTICS Press direct keys to ctrl FM See “Pump diagnostics” 4:21 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 95
Pump diagnostics The machine uses concentrate pump A and conductivity cell A to regulate the conductivity. In Pump Diagnostics the AK 95 S enters “nor- When the conductivity is stabilised (for 10 sec- mal” treatment mode, starts the fluid part and onds) inside ±5 % of the set value and the tem-...
Page 96
This mode can be used to check how many steps the stepping motor needs to move to obtain a flow of 500 ml/min. Start the AK 95 S, enter the Service menu and select Calibrations. Place A-pickup tube in the concentrate container.
Page 97
EVACUATION VALVE EVVA BYPASS VALVE CPU C BYVA BYPASS VALVE CPU P BYVA Note! Be careful when open the AIVA to avoid Fluid leakage into the monitor. 4:24 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 98
The actual information is shown on the informa- tion display, which is divided into three fields (0 - 2). It is possible to choose any AK 95 S parameter for any of the three fields. See Logging mode flow diagram.
Page 99
"Arrays", page 4:28 Select array index (000 - 999). Select the array to be displayed (see "Arrays", page 4:27) In which field the logged parameter will appear. 4:26 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 100
C COFB 088 043 20 Oct 05 16:37 For more information about the error message refer to page 4:6 and the Error Code List for AK 95 S, HCEN 9091. 4:27 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Page 101
Summary: 1. Set field 2. Set array 3. Set index 4. Push twice to step back 5. Push “display” to see the logged parameters. 4:28 AK 95 / AK 95 S Service Manual HCEN 9136 Rev. 10.2005...
Maintenance Manual AK 95 / AK 95 S October 2006 G A M B R O...
Page 104
Preventive Maintenance: B- and C-kit ... 5:17 Parts .............. 5:17 How to exchange the parts included in the B- and C-kit ......... 5:18 Actions to carry out after the parts have been exchanged ..........5:20 AK 95 / AK 95 S HCEN12189 Rev. 10. 2006...
Any warranties made by Gambro with respect to During repair of any of the parts in the flow the AK 95 / AK 95 S are void if the equipment is path, special care should be taken and a good not used in accordance with the instructions hygiene should be kept.
Page 106
Gambro K22151001 *BPM Cuff Gambro 100900141 *pH buffers ± 0.07 pH units pH 7 – 100810001 pH 9 – 100810002 Magnet for blood pump cover Gambro K19049001 * If installed AK 95 / AK 95 S HCEN12189 Rev. 10. 2006...
Page 107
Base-kit Preventive Maintenance: Base-kit Parts The Base-kit (K40182001) includes all parts necessary to fulfill the mandatory preventive maintenance procedures for AK 95 / AK 95 S. l i f l i f l i f r t l t l i...
Page 108
Then perform a heat disinfection program in combination with CleanCart A or a Chemical disinfection program with sodium hypochlorite. Refer to the AK 95 or AK 95 S Operator’s Manual. Change o-rings, 100319061, on the A and B-concentrate connectors (male).
Page 109
Maintenance Manual Change the O-rings, 100319029, and Change the level detector, K40136001, in square-ring, K14920001 in the ultrafilter the deairating chamber. Use the attached o-ring, 100319054. holder. (Only for AK 95 S with UFD-kit) 100319029 K40136001 K14920001 100319054 100319029 Change the sample port for 6 mm tube, Change both upper, 100319008, and lower K20219002.
Page 110
Change both upper, 100319008, and lower o-rings, 100319050, in the mixing chambers. 100319008 100319050 Mark month and year for the next scheduled preventive maintenance. Attach the label to the rear cover of the machine. AK 95 / AK 95 S HCEN12189 Rev. 10. 2006...
Page 111
Attach BPM cuff, 100900141, tight to the for AK-machines. test equipment, K22151001. Check the protective earth resistance: Start the AK 95 S with the main switch and the ON/OFF button. Measure between TP 5 (Z0VL) on the Enter Service mode.
Page 112
AK 95 S. In this case replace the cuff and cuff hose by a tube, which is closed. Note! Pressures shall be applied very carefully so that no overpressures occur.
Page 113
Maintenance Manual Base-kit Priming detector calibration Use following formula to calculate which Start the AK 95 S, enter the Service menu gage pin to use: and select Calibrations. 2 * wall thickness * 0,7 (mm) Select Transducer, External and PRIMING DETECTOR from the selection diagram.
Page 114
Calibrate if necessary (see chapter Transducer. calibration for details). Tolerance: ± 5 mmHg. Calibrate if necessary (see chapter Drain the AK 95 S and replace the blood calibration for details). leak calibration cover. Enter Diagnose/BM/Pumps and check that Attach pressure reference instrument the blood pump stops when the cover is before the heater.
Page 115
Check the DC leakage between TP 5 Turn on the power supply main switch (Z0VL) on the Control CPU board and the again and start the AK 95 S. suction pump head. Measure in DC Voltage, should be < 1 V.
Page 116
Simulate blood in the priming detector with a Perform heat/chemical disinfection. piece of paper. Make sure (again) that the measuring glass is filled to 1.00 liter. Press the UF START/STOP button. 5:13 AK 95 / AK 95 S HCEN12189 Rev. 10. 2006...
Page 117
Preventive Maintenance: Supplementary A-kit Parts The A-kit is designed to be used as a complemen- tary addition to the Base-kit for the preventive maintenance procedures of AK 95 / AK 95 S. The A-kit (K40179001) includes following exterior parts: e t l...
Page 118
K 40183001 (blue) and K 40184001 (red). - Pick-up tube, A (white) 100850007 - Pick-up tube, B (blue) 100850008 - Pick-up tube, (yellow) 100850021 A c e ta te K40183001 K40184001 5:15 AK 95 / AK 95 S HCEN12189 Rev. 10. 2006...
Page 119
Check the range between the blood First perform a heat disinfection program. pump rollers and the path according to Refer to the AK 95 or AK 95 S Operator’s the picture below. Manual. After the heat disinfection; re-tighten the nuts of the dialyser connector set to the machine.
Page 120
) . h , t i t r a o l ( , t i 5:17 AK 95 / AK 95 S HCEN12189 Rev. 10. 2006...
Page 121
B-kit only! Change all the Santoprene tubes in the Change valve membranes, K08056C, fluid monitor. for following valves: AIVA, ZEVA, DIVA, BYVA, TAVA, EVVA and FLVA. See instructions on page 5:20. 5:18 AK 95 / AK 95 S HCEN12189 Rev.10. 2006...
Page 122
Assemble according to machine configuration. Change the o-rings, 100335001, in the parker couplings to and from INPS. 100335001 100335001 Change the o-rings, 100335001, in the parker couplings from the heat exchanger. 5:19 AK 95 / AK 95 S HCEN12189 Rev. 10. 2006...
Page 123
Valve mounting tool K15391C A special service tool for mounting of the valve membrane correctly in the valve housing is available. 5:20 AK 95 / AK 95 S HCEN12189 Rev.10. 2006...
Page 124
Maintenance Manual Actions to carry out after the parts have been exchanged First perform a heat disinfection program. Refer to the AK 95 or AK 95 S Operator’s Manual. Let the machine pass FCH. 5:21 AK 95 / AK 95 S HCEN12189 Rev.
Page 125
Maintenance Manual This page is intentionally left blank 5:22 AK 95 / AK 95 S HCEN12189 Rev.10. 2006...
A piece of blood tube used at the hospital. Blood leak detector Calibration cover K40169001 Pressure Reference instrument 300333001 Calibration kit K13983002 Special tool For silicone tube connectors K12755001 Calibration kit AK 95 / AK 95 S Service Manual HCEN 9138 Rev. 10.2005...
Page 130
Calibrations AD Converters Set the “AD REF VOLTAGE” on the Start the AK 95 S, enter the Service menu display to the value shown on the voltme- and select Calibrations. ter and press Choose AD-converter. When selecting “AD-Converter” from the display...
Calibrations Transducers Calibration of linear transducers • Select Internal and the AK 95 S main- Start the AK 95 S, enter the Service tains a flow during the calibration. menu and select Calibrations. Independently wich flow support is chosen Choose Transducers the Selection diagram turns up.
+500 Connect the Gambro reference instrument (together with a calibration tube set) to Set the “REF SET” value on the AK 95 S the venous pressure transducer on the to the reading on the reference instrument front of the AK 95 S, see fig. below.
+500 the arterial pressure transducer on the front of the AK 95 S, see fig. below. Set the “REF SET” value on the AK 95 S to the reading on the reference instrument Zeroset the reference instrument, wait for and press stable reading.
Priming Detector calibration Occlusion adjustment, blood pump/ infusion pump Start the AK 95 S, enter the Service menu and select Calibrations. Check surface on the roller at the pump rotor, if it is damaged exchange the unit. The roller unit...
Calibrations Blood Leak Detector calibration Make sure there is no fluid in the AK 95 S If a calibration error occurs: Monitor. If fluid in the monitor, press Rinse/Drain. • Check the position of the calibration filter and repeat from point 1.
STOP Set the temperature to 37,5°C (by pressing into the EEPROM. TEMP key) and let the AK 95 S and the refer- ence instrument stabilize for at least 30 min- All the transducers A, B and P are all calibrated utes.
Page 137
Close clamp A. Apply a pressure of approx. -500 mmHg, shown on the reference instrument. Close clamp B. Set the “REF SET” value on the AK 95 S to the reading on the reference instrument and press twice.
Page 138
Calibrations Degassing Pressure Transducer Calibration Disconnect tube no. 50 at the Expansion Set the “REF SET” value on the AK 95 S to chamber. Connect the Gambro reference the reading on the reference instrument instrument to tube no. 50 via the calibra- and press tion tube set.
Remove the pH sensor from the pH sensor housing by loosening the nut holding the pH sensor and rinse the sensor in distilled water. Check that the “REF SET” on the AK 95 S displays pH 7. 6:12 AK 95 / AK 95 S Service Manual...
Page 140
STOP water. into the EEPROM. Check that the “REF SET” on the AK 95 S Remount the pH-sensor into the pH sensor displays pH 7. housing. Put the pH sensor in a cup with pH 7 If an error message is shown after this buffer solution (ordering no.
Page 141
PR 1 switch °C AIVA Suction Heat Pump exchanger Drain Outlet Restrictor AK 95 S Inlet/Outlet Flow Diagram. The complete Flow Diagram is showed at page 2:5. 6:14 AK 95 / AK 95 S Service Manual HCEN 9138 Rev. 10.2005...
Close clamp A. Apply a pressure of approx. -500 mmHg, shown on the reference instrument. Close clamp B. Set the “REF SET” value on the AK 95 S to the reading on the reference instrument and press twice.
This value will then be used during functional check and when entering service mode as the flow always should be 500 ml/min in functional check and service mode. Start the AK 95 S, enter the Service menu and select Calibrations. Place A-pickup tube in the concentrate container.
500 ml/min. It requires that the the UF-measuring unit. external UF calibration equipment is connected to the AK 95 S, see fig.below. The equipment The UF-measuring unit consists of two flow contains a reference volume, a flow directing measurement channels called channel 1 (CH 1) valve (UFVA) and a UF- calibration box.
Page 145
First the display for UF Volume for 3 seconds: (but nothing will get destroyed) if the cable is reversed. UF VOL REF VOL VOL DIFF 2. Start the AK 95 S with the main switch and 1006.5 1007.3 -0.28 % the ON/OFF button.
Page 146
Calibrations Intern When the internal UF calibration is ready: 1. Start the AK 95 S with the main switch and After the calibration , A K 9 5 S d i s p l a y s : the ON/OFF button.
Page 147
1006.5 1007.3 -0.28 % will get destroyed) if the cable is reversed. 2. Start the AK 95 S with the main switch and Then the display for Isolated UF Volume for 3 the ON/OFF button. seconds: 3. Place B-pickup tube in the acetate-concentrate ISO VOL REF VOL VOL DIFF container.
250 ± 3 mmHg 1. Attach cuff no. 3 tight to the BPM test equip ment, according to the illustrations below: 2. Start the AK 95 S with the main switch and the ON/OFF button. 3. Enter Service mode. 4. Enter Diagnostics mode.
Page 149
3 and cuff hose no. 1 to be connected to the check the actual pressure value for the BPM AK 95 S. Cuff no. 3 to be wrapped tight to the pressure on the AK 95 S display. BPM test equipment K2 2151 001.
Page 150
Possible causes are pressure transducer or electronic circuits mal- function. The BPM module must then be re- placed and sent to Gambro for repair. 6:23 AK 95 / AK 95 S Service Manual HCEN 9138 Rev. 10.2005...
Page 151
Calibrations This page is intentionally left blank 6:24 AK 95 / AK 95 S Service Manual HCEN 9138 Rev. 10.2005...
Page 154
Spare Part List Item Spare Part List Denomination Order no. Remarks 10.2005 Contents : Front panel outside ........ 7-1 Front panel inside ........7-2 Operators panel ........7-3 Blood pump ..........7-4 Blood pump, Motor unit ......7-5 Heparin pump ........7-6 Bottom box ..........
7-1 Front panel outside 5,24 HCEN9094 Rev. 10.2005...
Page 156
Spare Part List Item Denomination Order no. Remarks Front panel K21747001 AK 95: K21747002 O-ring 100318054 100320065 Cabling, reciving K21166001 O-ring 100318064 PT-screw 100366314 Level detector door K15114002 Shaft K17529001 Spring 100346026 Roller K12994001 Lundia 11 arm K15541B Multi holder K20397003 Lundia Multi holder N01870003...
Page 190
Spare Part List Item Denomination Order no. Remarks Protective Driver board K17552003 with frame Protective CPU board K17553005 with frame Control CPU board K17554006 with frame Control Driver board K17555004 with frame Quick-acting lock K04813001 PT-screw 100368410 Screw 100370440 Screw 100370425 Fiber washer 100322049...
Page 192
Spare Part List Item Denomination Order no. Remarks Infusion stand,straight K19073B Screw 100388612 Knob 100300048 Cover, rear K18493B AK 95: K18493A Air filter K12966001 Rear cover Infusion stand,angled K19073A Filter holder K16963001 Insert holder, left K12731A Included in item 4 Insert holder, right K12963A Included in item 4...
7- 21 MM95 with Stand 24,25 HCEN9094 Rev. 10.2005...
Page 198
Spare Part List Item Denomination Order no. Remarks Screw 100370410 Front K19840D AK 95: K19840C Stand K16272B Cover, left K16494D AK 95: K16494C Cover, right K16495004 AK 95: K16495003 Screw 100378406 Locking fluid Screw 100370440 Frame, upper complete K16703D AK 95: K16703C Screw 100388812 Plug...
Page 220
Spare Part List Item Denomination Order no. Remarks 1 Protective Driver Board K17552003 ( with frame ) P200 cabling K18133001 to control Driver board P408 P201 cabling K21166001 to air detector P202 cabling K18169001 to FM I/O board P509 P203 to protection CPU board P100 P204 cabling...
Page 222
Spare Part List Item Denomination Order no. Remarks 6 MM I / O Board K40090002 P700 to pH-electrode P701 cabling K11291001 to conductivity cell A P702 cabling K11291001 to conductivity cell B P703 cabling K18205001 to BiCart holder and pick-up tubes P704 cabling K16727001...
Chapter 9 Symbol Description HCEN9160 Revision 10. 2005 Service Manual - Symbols...
Page 226
List of symbols NOTE All symbols in this list may not be represented on this product. Description Alternating current Protective earth (ground) Warning, consult accompanying documents Off (power, disconnection from the mains) On (power, connection to the mains) Type B, applied part Handle with care This way up Keep dry...
Page 227
CSA approved for Canada and USA CSA-listed label manufacturer Prohibited behaviour Equipotentiality Separate collection for electrical and electronic equipment HCEN9160 Revision 10.2005 Service Manual - Symbols...
Page 228
This page is left intentionally blank HCEN9160 Revision 10. 2005 Service Manual - Symbols...
Replacements Replacements Replacement matrix........2 Changeactions ............. 4 FM Flow Diagram ..........5 10:1 AK 95 / AK 95 S HCEN12078 Rev.10.2005...
Page 230
In the matrix, BM stands for Blood Monitor and FM stands for Fluid Monitor. The replacementmatrix is only a guideline for service technicians which actions to perform when a component has been replaced. 10:2 AK 95 / AK 95 S HCEN12078 Rev.10.2005...
Page 231
Replacements 10:3 AK 95 / AK 95 S HCEN12078 Rev.10.2005...
Page 232
Calibrate according to calibration instructions in the Service Manual, Calibration Manual. the Service Manual, Calibration Manual. Start the machine and let it perform a Function Check. Heat disinfection Perform a heating according to the Operators Manual. 10:4 AK 95 / AK 95 S HCEN12078 Rev.10.2005...
Page 233
Replacements FM Flow Diagram 10:5 AK 95 / AK 95 S HCEN12078 Rev.10.2005...
Page 234
Replacements This page is intentionally left blank 10:6 AK 95 / AK 95 S HCEN12078 Rev.10.2005...
Electrical Safety Inspections for AK-machines Including following sub-tests: - Visual inspection - PET - Protective earth test - ELT - Earth leakage current test - PLT - Patient leakage current test Electrical Safety Inspections for AK-machines 11:1 HCEN12190 Rev 05.2006...
Page 236
General information To ensure proper operation, a qualified service technician must per- form an Electrical Safety Inspection (ESI) of the AK-machine at reg- ular intervals. The information needed to perform ESI is provided in this manual. Included in the ESI procedures are checks to verify normal machine operation.
Page 237
Visual inspection The visual inspection is an important part of the electrical safety inspections of the AK-machine. The visual inspection is a procedure to make sure that the medical equipment in use still confirms the specifications and has not suffered from any external damage and/or contamination.
Page 238
Manufacturer: e.g. Fluke ± 0.5 % Resistance: accuracy PET for AK 95 / AK 95 S Check that the resistance between the grounding clips of the mains plug and the exterior cover of the power supply is < 0.3 Ohm.
Page 239
PET for AK 200 / AK 200 ULTRA / AK 200 S / AK 200 ULTRA S Check that the resistance between the grounding clips of the mains plug and the power supply (right side, through the vent- holes of the bottom box) is < 0.3 Ohm. Check that the resistance between the grounding clips of the mains plug and the FM bottom plate is <...
Page 240
ELT / PLT Test equipment Safety tester: According to IEC 60601-1 Suggested Manufacturer: Metron, Rigel, Biotec General conditions for ELT / PLT The AK-monitor shall be connected to both feed water and drain. Connect the AK-monitor to the outlet supply of the safety tester. Use a mains plug adapter for respective mains plug.
Page 241
Take the highest reading when the machine is running. Check that the highest measured readings don’t exceed the limit values in following table: Limit values for ELT AK 95 S (100V) Max 500 µA Typical < 100 µA AK 95 S (110V) Max 500 µA...
Page 242
Take the highest DC-reading when the machine is running. Check that the highest measured readings don’t exceed the limit values in following table: Limit values for PLT AK 95 S (100V) Max 100µA Max 10µA Typical <10µA AK 95 S (110V) Max 100µA...
Page 243
NOTE For treatments with a central venous catheter (CVC), the patient leakage current must be < 10µA Electrical Safety Inspections for AK-machines 11:9 HCEN12190 Rev 05.2006...
Page 244
This page is intentionally left blank 11:10 Electrical Safety Inspections for AK-machines HCEN12190 Rev 05.2006...
Page 245
Visual inspection Visual inspection of the AK-machine according to section 2. PET for AK 95, AK 95 S (1) Resistance between the grounding clips of the mains plug and the exterior cover of the power supply is < 0.3 Ohm.
Page 246
ELT - Earth Leakage Current Test File the highest earth leakage current value when the machine is running according to section 4.3: Earth leakage current value: _______ µA Note! Check that the highest measured readings don’t exceed the limit values described in section 4:3 PLT - Patient Earth Leakage Current Test File both the highest AC-reading and the highest DC-reading when the...