General Alarms and Troubleshooting

AC PUMP ERROR!

AC PUMP ERROR! Check for proper
loading or leaks.
Alarm LED is red. Pumps stop.
Probable Cause
Tubing incorrectly positioned in the AC pump, valves, or
access pressure sensor
Leak at the access pressure sensor pod seal
Leak in the disposable tubing set
Pump, valve, or sensor malfunction
12-10
CONTINUE
Ensure the tubing is placed correctly in the AC pump, valves, and access pressure sensor.
Press CONTINUE.
Check the access pressure sensor pod seal for wear or damage. Replace if necessary. For instructions, see
"Changing pressure pod seals" on page 10-4.
Check the disposable tubing set for leaks. If the set does not leak, press CONTINUE. If the set leaks, replace it
with a new set and prime the set.
If the alarm recurs, replace the disposable tubing set.
Note the alarm message. Discontinue the procedure. Notify a qualified service technician of the alarm
message.
To retry the AC pump test, press CONTINUE.
This alarm occurs only during Prime mode, when the Spectra system
tests the AC pump. It occurs when the AC pump fails to develop
sufficient pressure as measured by the access pressure sensor.
.
Action
Note: If the access pressure sensor is unloaded while there is fluid in the disposable tubing set,
see "Repositioning the Access Pressure Sensor Diaphragm" on page 11-9 for instructions on
repositioning the diaphragm and reloading the access pressure sensor.
Note: For instructions on repositioning the diaphragm and reloading the access pressure sensor,
see "Repositioning the Access Pressure Sensor Diaphragm" on page 11-9.
®