General Alarms and Troubleshooting

CCM not operational - this run only.

CCM not operational – this run only.
Warning and Alarm LEDs are off.
Probable Cause
Saline calibration failed at end of Prime mode due to a
problem with cuvette installation (cuvette contained bubbles
or it was installed incorrectly)
CCM malfunction
12-20
This message appears only if the CCM display is selected during the Run
mode and saline calibration (during Prime mode) was not successful. A
CCM CALIBRATION FAILURE alarm should have occurred during
Prime mode.
To remove the message, press CLEAR. The CCM is not accurate during this Run mode only. Monitor the
channel for RBC accumulation. During AutoPBSC procedures, open the collect valve during the Harvest
phase (see "Assisted Mode" in Chapter 4 of the Cell Therapy Guide).
If the message continues to appear, notify a qualified service technician.
Action
®
COBE
Spectra™ Apheresis System • Essentials Guide