Waste Divert Valve
Press VALVE, then press 3 to select "Waste." The following screen
appears:
Waste: 1=Inlet Divert, 2=Return Divert,
3=Close, 4=Recirculate, 5=Load.
See Figure 5-1 for a diagram of the waste divert valve positions.
Use Table 5-2 to help you decide which number to press. For example, if
you press 5, the waste divert valve moves to the load position, and all
three lines (waste, inlet waste, and return waste) open.
Waste Line
Inlet waste line
®
COBE
Spectra™ Apheresis System • Essentials Guide
Inlet divert
Return divert
Return waste line
Figure 5-1: Waste Divert Valve positions
Table 5-2: Waste divert valve positions
Press...
Waste Line
Inlet Waste Line
1
Open
2
Open
3
Closed
4
Closed
5
Open
Close
Recirculate
Software Options and Configuration
Return Waste Line
Open
Closed
Closed
Open
Open
Closed
Open
Open
Open
Open
Load