Pumps Off Too Long. Centrifuge Will Turn Off In __ Seconds; Pumps Off Too Long. Centrifuge Turned Off For Donor/Patient - Gambro Cobe Spectra Essentials Manual

Table of Contents

Advertisement

General Alarms and Troubleshooting

PUMPS OFF TOO LONG. Centrifuge will turn off in __ seconds.

PUMPS OFF TOO LONG. Centrifuge will
turn off in ____ seconds.
Alarm LED is red. Pumps stop.
Probable Cause
Pumps paused for more than 9 minutes
PUMPS OFF TOO LONG. Centrifuge turned off for donor/patient safety.
PUMPS OFF TOO LONG! Centrifuge turned
off for patient safety.
Alarm LED is red. Pumps and centrifuge stop.
Probable Cause
Pumps paused for more than 10 minutes
12-50
CONTINUE*
If the pumps are to remain paused, continue a saline drip. The centrifuge will turn off after 60 seconds. To
restart the pumps when ready to continue the procedure, press CONTINUE.
CONTINUE*
If the pumps are to remain paused and the centrifuge is turned off, continue a saline drip. To restart the pumps
and centrifuge when ready to continue the procedure, press CONTINUE.
To override the alarm, press CONTINUE.
This alarm occurs only during Run mode.
When this alarm occurs, the screen begins to count down 60 seconds.
For donor/patient safety, the centrifuge stops when the countdown
reaches zero, and the "PUMPS OFF TOO LONG" alarm occurs.
Action
To override the alarm, press CONTINUE.
This alarm occurs only during Run mode.
Action
®
COBE
Spectra™ Apheresis System • Essentials Guide