Resolving Power-Up Tests Failure; Repositioning The Access Pressure Sensor Diaphragm - Gambro Cobe Spectra Essentials Manual

Table of Contents

Advertisement

9
Turn the plasma pump counterclockwise approximately
100 revolutions.
10
Close the white pinch clamp on the return line, then disconnect the
return needle.
11
Close the access saline line roller clamp.
12
Remove the disposable tubing set from the front panel:
a
Unlock the centrifuge cover (see "Manually Overriding the
Centrifuge Cover and Door Latches" on page 11-7), then
remove the channel from the filler.
b
Turn the valves to the open/load positions.
c
Press the cartridge clamps to release the pump cartridges. Turn
the pumps counterclockwise and pull the tubing out to unload
the pumps.
d
Remove the lines from the front panel.
e
Discard the containers, bags, and disposable tubing set in the
biohazard disposal container according to your standard
operating procedures.

Resolving Power-Up Tests Failure

If the "Power up tests in progress" appears for more than 60 seconds,
perform one or both of the following:
1
Turn the power switch off and then on again to retry the power-up
tests.
2
Try another power outlet or circuit in case other equipment is
reducing the power available to the Spectra system.
If the problem persists, notify a qualified service technician or
Gambro BCT.
®
COBE
Spectra™ Apheresis System • Essentials Guide
Repositioning the Access Pressure Sensor
Diaphragm
Repositioning the diaphragm with fluid in the disposable tubing set
during Prime or Run mode may be necessary when
The access pressure sensor has not been positioned correctly or has
been disengaged from the access pressure sensor housing.
The diaphragm is not in the proper position because the pressure
pod seal is dry or cracked.
To reposition the diaphragm in the access pressure sensor so the
transducer in the housing can monitor access pressure accurately,
perform the following:.
Note: Before repositioning the diaphragm, pause the system.
1
Remove the access pressure sensor from its housing. Turn
counterclockwise to unlock.
2
If in Run mode, close the access line pinch clamp and open the
access saline roller clamp.
3
Lower the saline container until the fluid level in the container is
level with the centrifuge cover (see Figure 11-2). Wait
approximately 5 seconds.
4
Raise the saline container until fluid in the container is level with
the pressure sensor (see Figure 11-2).
5
Wait approximately 10 seconds.
6
Load the pressure sensor into its housing, keeping the fluid level in
the saline container level with the sensor. Push in and turn the
sensor clockwise to lock.
7
Rehang the saline container.
Helpful Hints and Recovery Procedures
11-9