Software Options and Configuration
AC Data
To display the AC data screen, press 4. The following screen appears:
AC infusion rate: ___ ml/min/liter TBV.
ml AC in bags: collect:___, plasma:__
This screen displays the current AC infusion rate and the predicted
volume of AC in the collection and plasma bag(s) at the end of the Run.
Note: This screen is not available during RBCX or BMP
procedures.
Option 2: Pressure Display
Press MENU ON/OFF, then press 2 to select "Pressure Display." The
following screen appears:
1 = Current pressures,
2 = Set alarm limits.
To display the current access and return pressures, press 1.
To change the access and return pressure alarm limits during the
Run mode, press 2.
5-10
Current Pressures
___.__ ___.__ ___.__ ___.__ ___.__ _____
Access: ____, Return: ____ (mmHg)
Use this screen to monitor the access and return pressures. (During
Single-Needle procedures, only the access pressure appears.) The
pressures appear on the bottom line and the pump flow rates appear on
the top line, allowing you to adjust the pump flow rates and observe the
changes in the access and return pressures at the same time.
Set Alarm Limits
You cannot view this screen before starting the Run mode. If you are in
Prime mode when you select this option, the following screen appears:
Must be in Run before alarm limits
can be changed!
If you are in the Run mode when you select this option, the following
screen appears:
Access pressure low limit:
Return pressure high limit:
Note: You cannot adjust the return pressure high limit during
Single-Needle procedures.
®
COBE
Spectra™ Apheresis System • Essentials Guide
–{250}
400