AC Power
Characteristics
Line voltage
Input line current
Electrical Safety
Characteristics
AC leakage current
Safety Certifications
Characteristics
Certified by CSA International to CAN/CSA 60601-1 and
UL 60601-1
CE-Marked under the Medical Device Directive
®
COBE
Spectra™ Apheresis System • Essentials Guide
Performance
100 ±10% volts A.C., 47–63 Hz
115 ±10% volts A.C., 47–63 Hz
220 ±10% volts A.C., 47–63 Hz
240 ±10% volts A.C., 47–63 Hz
8 A maximum RMS at 100/115 volts A.C.
4 A maximum RMS at 220/240 volts A.C.
50 microamperes maximum RMS, 115 volts A.C.
100 microamperes maximum RMS, 100/220/240 volts A.C.
Performance
Bears CSA marking with adjacent "C" and "US" for
Class 1, Type BF Equipment.
Certified file no. LR 114324-1
Notified body: British Standards Institution
Conditions
Circuit breaker overcurrent protection
Performance
Conditions
115 volt units in North America
High voltage units