Warning: The ELP disposable tubing set is no longer functionally
closed if
•
The first attempt to successfully insert the access needle fails
(unless you attach a new needle using an approved
connection device which maintains the sterility of the
disposable tubing set).
•
The access needle (or, in the Single-Needle procedure, the
access/return needle) is disconnected during the procedure.
•
During a Single-Needle procedure, you use the injection site
on the access manifold to obtain blood samples or to infuse
medication or parenteral solutions.
•
You remove samples before you seal the tubing to the
sampling bag.
•
You disconnect any product bag before you seal it.
•
The integrity of the disposable tubing set is compromised for
any reason.
If any of these occur, DO NOT store the product for more than
24 hours.
®
COBE
Spectra™ Apheresis System • Essentials Guide
Warning: The LRS disposable tubing set is no longer functionally
closed if
•
The first attempt to successfully insert the access needle fails
(unless you attach a new needle using an approved
connection device which maintains the sterility of the
disposable tubing set).
•
The access needle (or, in the Single-Needle procedure, the
access/return needle) is disconnected during the procedure.
•
During a Single-Needle procedure, you use the injection site
on the access manifold to obtain blood samples or to infuse
medication or parenteral solutions.
•
You remove samples before you seal the tubing to the
sampling bag.
•
You disconnect any product bag before you seal it.
•
The integrity of the disposable tubing set is compromised for
any reason.
If any of these occur, DO NOT store the product for more than
24 hours.
Warning: Adding calcium gluconate or other calcium salts to
fresh frozen plasma can cause clotting in the replacement fluid.
Blood products containing formed elements are not recommended
replacement fluids.