General Alarms and Troubleshooting
Probable Cause (continued)
Leak at return pressure sensor pod seal
Internal system malfunction

Centrifuge up to speed.

__.__
___.__ ___.__ ___.__ ___.__ ____
Centrifuge up to speed.
Warning LED is yellow.
Probable Cause
This warning occurs after you press CONTINUE to restart
the centrifuge (after it had been stopped by an alarm or by
pressing STOP SPIN) and the centrifuge is up to speed
again
12-24
Check the return pressure sensor pod seal for wear or damage. Replace if necessary. For instructions, see
"Changing pressure pod seals" on page 10-4.
Note the alarm message. Press CONTINUE.
If the alarm recurs, discontinue the procedure. Press CHANGE MODE, then press 4 to select "Rinseback." To
override the alarm during Rinseback mode, press CONTINUE. Notify a qualified service technician of the alarm
message.
CONTINUE
To remove the warning message and start the pumps, press CONTINUE.
Action (continued)
Note: For instructions on repositioning the diaphragm and reloading the return pressure sensor,
see "Repositioning the Return Pressure Sensor Diaphragm" on page 11-10.
Action
®
COBE
Spectra™ Apheresis System • Essentials Guide