Alarm Test Failed; Ccm Calibration Failure - Gambro Cobe Spectra Essentials Manual

Table of Contents

Advertisement

ALARM TEST FAILED!

ALARM TEST FAILED!
Press CONTINUE to retry alarm tests.
Warning and Alarm LEDs are off. Pumps and centrifuge stop.
Probable Cause
An alarm test failed
Ensure the disposable tubing set is properly loaded and that the centrifuge door is completely closed. To retry the alarm tests, press
CONTINUE.
Internal system malfunction
Note the message. Discontinue the procedure. Notify a qualified service technician of the message.

CCM CALIBRATION FAILURE.

CCM CALIBRATION FAILURE.
Press 1 to retry, or CLEAR to disable.
Warning LED is yellow.
Probable Cause
Cuvette incorrectly positioned in the CCM
Bubble in cuvette
CCM malfunction
®
COBE
Spectra™ Apheresis System • Essentials Guide
.
Action
To remove the warning message and disable the CCM during this Run
mode, press CLEAR.
This warning occurs at the end of Prime mode if CCM calibration was
not successful.
.
Ensure the cuvette is placed correctly in the CCM. To retry CCM calibration, press 1.
Tap the cuvette to move the bubble. To retry CCM calibration, press 1.
Note the warning message. Notify a qualified service technician of the warning message.
General Alarms and Troubleshooting
Action
12-19