Specifications
Sensors
Access Pressure Sensor
Characteristics
Operating range
Default alarm pressure
Operator-adjusted alarm pressure
Accuracy
Return Pressure Sensor
Characteristics
Operating range
Default alarm pressure
Operator-adjusted alarm pressure
Accuracy
Centrifuge Pressure Sensor
Characteristics
Operating range
Alarm pressure
13-8
Performance
–300 to +50 mm Hg
–250 ±30 mm Hg and lower default
–50 to –250 ±30 mm Hg
±12% of reading or ±20 mm Hg, whichever is greater
Performance
–50 to 450 mm Hg
Single-Needle ELP procedure: –50 to +500 mm Hg
400 ±48 mm Hg
Single-Needle ELP procedure: 450 ±48 mm Hg
50 to 400 ±48 mm Hg
±12% of reading or ±20 mm Hg, whichever is greater
400 to 1,200 mm Hg
1,000 mm Hg
Conditions
Dual-needle and Single-Needle ELP procedures: Alarm
pressure during Rinseback mode is –300 mm Hg
Current Run mode only
Conditions
Current Run mode only
Single-Needle procedure: Run mode is not adjustable
Performance
®
COBE
Spectra™ Apheresis System • Essentials Guide