Gambro Cobe Spectra Essentials Manual page 218

Table of Contents

Advertisement

The Seal Safe
System
Replacing a Splash Guard
1
Disconnect the cutter/sealer head from the RF interconnect cable
by disconnecting the RF cable at the cutter/sealer head.
2
Remove and discard the existing splash guard.
3
Remove the fully perforated clip on one side of a new splash guard.
Fold the other clip inward at a 90° angle (see Figure 15-8).
Figure 15-8: Remvoing and folding the clip of a new splash guard
4
Fold the sides of the splash guard inward at the perforations (see
Figure 15-9).
Figure 15-9: Folding the sides and clip of a new splash guard
15-6
5
Align and clip the new splash guard to the lever assembly (see
Figure 15-10).
Figure 15-10: Aligning the new splash guard
6
Bend the end of the clip and lock the splash guard in place (see
Figure 15-11).
Figure 15-11: Locking the splash guard in place
®
COBE
Spectra™ Apheresis System • Essentials Guide