General Alarms and Troubleshooting

FAILURE #_: __ __ __ _ _

FAILURE #__: _____ _____ _____ __ __
Record #.CONTINUE to retry or Rinseback.
Alarm LED is red. Pumps stop. Depending on alarm, the centrifuge
stops and/or the return line valve closes.
Probable Cause
Internal system malfunction
12-34
This alarm does not include "FAILURE #3: __Display Comm__ __ _"
or "FAILURE #20: ___ ___ ___ _ _".
Record the entire alarm message. The code and text data it contains could help a qualified service technician
diagnose the cause of the problem.
Press CONTINUE. If the alarm clears and does not recur, continue the procedure.
If the alarm does not clear or continues to occur, discontinue the procedure. Press CHANGE MODE, then press
4 to select "Rinseback." To override the alarm during Rinseback mode, press CONTINUE. Notify a qualified
service technician of the alarm message.
If you cannot override the alarm during Rinseback mode, turn off the system. See the "Manual Rinseback
Procedure" on page 11-8.
Action
®
COBE
Spectra™ Apheresis System • Essentials Guide