Mains Voltage and Power
Consumption
The AK 95 S must be plugged into a grounded
power point not more than 3 metres distant.
Mains voltage:
100 V(at 50 Hz or 60 Hz)
110 V(at 50 Hz or 60 Hz)
230 V(at 50 Hz or 60 Hz) (±12%)
Power consumption:
2025 W at 230 V
1575 W at 110 V
1275 W at 100 V
Service Outlets
One extra grounded power socket should be
available close to the AK 95 S, for troubleshoot-
ing and maintenance purposes.
Line Fuses
Never overload the line fuses and do not supply
too many systems via a single line fuse.
Mains Water Supply
Required quality:
Level of purity according to ASAIO and AAMI
recommended standards on feed water for
dialysis monitors.
Water for the dialysis unit has to be supplied
from a Reverse Osmosis purification system.
Pressure:
1.2 - 6 bar to the pressure regulator mounted in
the MM 95.
Note! The pressure should be measured with
the flow from the water supply.
Temperature:+5 °C to +30 °C (treatment)
Consumption in treatment:
AK 95 S: 500 ml/min
AK 95: 550 ml/min
Drain: Open type. Not more than 1.2 metres
above outlet from fluid monitor.
Note! Local ordanances may require the use of
special measures to protect against the possibil-
ity of back-siphonage from dialysis equipment
into the water supply. When installing the
AK 95 S, such ordanance should be observed.
AK 95 / AK 95 S Service Manual
HCEN 9133 Rev. 01. 2004
Tools
The tools listed below are needed during the
installation procedure.
•
Tube cutter, 113 500 084
•
Adjustable wrench, 113 003 009
•
Hexagon spanner set
•
Screwdriver, 1,2 x 8,0 - 175mm,
113 002 006
Top Tray
To avoid leakage into the machine, the top tray
must always be placed on the top of the ma-
chine, except during technical service.
Lift points
Installation Guide
B
A
Acet ate
Lift points
Figure 2. AK 95 S
1:3