Extracorporeal Volume - Gambro Cobe Spectra Essentials Manual

Table of Contents

Advertisement

Table 9-2: Volume returned to patient at each state of Rinseback
Procedure
ELP (Dual-Needle and Single-Needle)
LRS and LRST (Dual-Needle and Single-Needle)
WBC procedures
AutoPBSC
*
TPE (Dual- and Single-Needle), RBCX
* Rinseback is not recommended for RBCX procedures.

Extracorporeal Volume

The tubing and separation channel form the extracorporeal circuit to and
from the donor/patient. The tubing configuration minimizes extracorporeal
volume. Each disposable tubing set holds less than 285 mL of whole blood
equivalent. The AutoPBSC disposable tubing set holds an extracorporeal
volume of less than 165 mL, making it suitable for use with pediatric
patients.
®
COBE
Spectra™ Apheresis System • Essentials Guide
Collect
Returning RBCs
60 mL
90 mL
N.A.
160 mL
60 mL
120 mL
N.A.
175 mL
N.A.
120 mL
Rinseback States
Evacuating Channel
Rinsing Channel
100 mL
100 mL
125 mL
100 mL
150 mL
Fluid Volumes
TOTAL
60 mL
310 mL
60 mL
320 mL
108 mL
413 mL
60 mL
335 mL
75 mL
345 mL
9-5