Donor/Patient Disconnected???

Donor/Patient Disconnected???
PUMPS WILL UNLOAD!!!
Alarm LED is red.
Probable Cause
Access, AC or access saline line not clamped
– or –
Access pressure sensor removed from housing before
unloading the pumps
– or –
Access pressure sensor malfunction

Ensure return flow scale is in prime position.

Ensure return flow scale
is in prime position.
Alarm LED is red. Pumps stop.
Probable Cause
Return Flow Controller not in the prime position
Waste valve not operating correctly after prime divert
®
COBE
Spectra™ Apheresis System • Essentials Guide
(YES/NO)
Ensure the donor/patient has been disconnected. Clamp and disconnect access and return lines. Close both
access pinch clamps. Clamp and seal the AC and access saline lines.
To retry the donor disconnect test, press NO.
– or –
To unload the pumps, press YES.
If this alarm recurs, note the alarm message. Notify a qualified service technician of the alarm message.
CONTINUE
Place the Return Flow Controller in the prime position.
See "WASTE VALVE NOT OPERATING CORRECTLY!" on page 12-63.
To unload the pumps, press YES.
.
Action
This alarm occurs only during Single-Needle procedures when the run
time valve check fails.
Action
General Alarms and Troubleshooting