OVERTORQUE ON _______ PUMP!
OVERTORQUE ON ____ PUMP!
Check pump tubing.
Alarm LED is red. Pumps stop.
Probable Cause
Tubing jammed around the pump
Object in rotor raceway
Internal system malfunction
®
COBE
Spectra™ Apheresis System • Essentials Guide
CONTINUE
Ensure the tubing is properly loaded in the pumps.
Correct the tubing jam. (Unload and reload the pump cartridges, if necessary. Clamp the AC line before
unloading the pumps.) Turn the pump to ensure it moves freely.
Press CONTINUE.
Remove the object. Press CONTINUE.
Note the alarm message. Press CONTINUE.
If the alarm recurs, discontinue the procedure. See "Manual Rinseback Procedure" on page 11-8. Notify a
qualified service technician of the alarm message.
General Alarms and Troubleshooting
Warning: DO NOT unload pumps while the donor or patient is
connected.
Action
12-45