General Alarms and Troubleshooting
Probable Cause (continued)
Needle too small
– or –
Inlet pump flow rate too high
Defective return pressure sensor
– or –
Internal system malfunction
12-56
Lower the inlet pump flow rate.
Note the alarm message. Press CONTINUE.
If the alarm recurs, discontinue the procedure. Press CHANGE MODE, then press 4 to select "Rinseback." To
override the alarm during Rinseback mode, press CONTINUE. Notify a qualified service technician of the alarm
message.
Action (continued)
®
COBE
Spectra™ Apheresis System • Essentials Guide