Setting Up The Spectra System - Gambro Cobe Spectra Essentials Manual

Table of Contents

Advertisement

Loading and Removing the Disposable Tubing Set and Other Common Operations

Setting up the Spectra System

1
Plug in the Spectra system.
2
Turn the power switch to the on position.
The system performs a short self-check to ensure the power supplies
are operating at the correct voltages:
Power up tests in progress.
3
A screen similar to one of the following with the correct Spectra
version appears:
COBE Spectra (Version 4.7 Software).
Press CONTINUE to load tubing set.
COBE Spectra (Version 5.1 LRS Software).
Press CONTINUE to load tubing set.
COBE Spectra (Version 6.1 PBSC Software)
Press CONTINUE to load tubing set.
COBE Spectra (Version 7.0 LRS Turbo).
Press CONTINUE to load tubing set.
4
Verify the following:
Yellow Warning LED illuminates
Pause LED flashes
Cartridge clamps are in the unload position
4-2
Removing and Installing the Filler
Caution: Before installing a filler or loading a disposable tubing
set, inspect the filler and visible interior of the centrifuge chamber
for fluid spills or structural damage. Ensure the metal band around
the perimeter of the filler is centered and secure.
Table 4-1 shows the filler to use for each apheresis procedure. See the
procedural guides, or Figure 2-6 on page 2-17 for illustrations of the
fillers.
Table 4-1: Fillers to use for procedures
Procedure
BMP
LPE
MNC Collection
PMN Collection
RBCX
TPE
WBC Depletion
AutoPBSC Collection
*
Platelet Depletion
ELP Collection
LRS Collection
LRST Collection
* You may perform Platelet Depletion procedures with any dual-stage
filler, but you must use the correct disposable tubing set for the filler.
See Chapter 5, "Platelet Depletion Procedure," of the Therapeutic
Apheresis Guide for more information.
®
COBE
Spectra™ Apheresis System • Essentials Guide
Filler
Single-stage
Dual-stage
Dual-stage with LRS bracket
Dual-stage LRS Turbo