Net Additional Saline Returned To The Donor Or Patient - Gambro Cobe Spectra Essentials Manual

Table of Contents

Advertisement

Fluid Volumes
Net Additional Saline Returned to the Donor or
Patient
Use Table 9-3 to determine the net amount of additional saline returned to
a donor/patient at the end of Rinseback mode.
Table 9-3: Net amount of additional saline returned to donor/patient
Procedure
ELP
Platelet Depletion
LRS and LRST
LPE
MNC and PMN Collection
WBC Depletion (MNC and PMN)
AutoPBSC
TPE
*
RBCX
* Rinseback is not recommended for RBCX procedures.
9-6
Total Volume of Saline Processed
During Rinseback
+310 mL
+320 mL
+413 mL
+335 mL
+345 mL
+345 mL
.
Amount of Saline Diverted to
Net Additional Saline Returned to
Waste Bag During Waste Divert
–120 mL
–120 mL
–150 mL
–150 mL
–150 mL
–100 mL
COBE
Donor/Patient
+190 mL
+200 mL
+263 mL
+185 mL
+195 mL
+245 mL
®
Spectra™ Apheresis System • Essentials Guide