General Alarms and Troubleshooting

EXCESSIVE LOAD ON CENTRIFUGE!

EXCESSIVE LOAD ON CENTRIFUGE!
Check centrifuge disposables.
Alarm LED is red. Pumps and centrifuge stop.
Probable Cause
Channel or multi-lumen tubing improperly installed
Internal system malfunction
12-30
CONTINUE
Ensure the channel and multi-lumen tubing are properly installed in the centrifuge.
Note the alarm message. Press CONTINUE.
Important: Press CONTINUE only once, then call a qualified service technician. This alarm is often due to an
electrical failure, and pressing CONTINUE more than once can damage internal elements.
If the alarm recurs, discontinue the procedure. Press CHANGE MODE, then press 4 to select "Rinseback." To
override the alarm during Rinseback mode, press CONTINUE. Notify a qualified service technician of the alarm
message.
To override the alarm for 60 seconds, press CONTINUE.
Action
®
COBE
Spectra™ Apheresis System • Essentials Guide