Loading and Removing the Disposable Tubing Set and Other Common Operations
16
Remove the lines from the following:
Collect/replace and plasma valves
Return pressure sensor
Waste divert valve
RBC line valve
Collect concentration monitor (CCM) (if the disposable
tubing set includes a cuvette)
Return and inlet air detectors
Centrifuge pressure sensor
Access pressure sensor
Return line valve
AC level detector
17
Press the cartridge clamps to release the pump cartridges, and
remove the cartridges from the clamps.
18
Remove any needles still attached from the disposable tubing set,
including the needle in the saline container, and place them in an
appropriate needle disposal container.
19
Single-Needle procedures: Remove the Single-Needle bag from the
Return Flow Controller.
20
Discard the containers, bags, and disposable tubing set in a
biohazard disposal container according to your standard operating
procedures.