Ensure the correct filler is installed. If it is not, perform the following to
remove the existing filler and install the correct one. See Figure 2-4 on
page 2-15 for an illustration of the centrifuge chamber:
1
Press UNLOCK COVER.
2
Slide back the centrifuge cover and lower the centrifuge door.
3
Rotate the centrifuge so that the centrifuge loading port is facing
the front.
4
Remove a filler as follows:
a
Push the filler latching pin toward the center of the centrifuge
and raise the filler latch.
b
Push the filler locking pin toward the center of the centrifuge
and lift the filler. Remove it from the centrifuge chamber.
5
Position the new filler so that the dots on the centrifuge and filler
align (see Figure 4-1).
Dual-stage Turbo fillers do not have dots. Find the notch on the
bottom of the filler. Line up this notch with the pin on the
centrifuge.
6
Place the filler over the centrifuge assembly, and press down until
the filler locking pin is securely in place.
7
Lower the filler latch.
8
Lift up on the filler to ensure it is securely in place.
9
Close the centrifuge door and cover.
Continue with "Loading the Disposable Tubing Set."
®
COBE
Spectra™ Apheresis System • Essentials Guide
Loading and Removing the Disposable Tubing Set and Other Common Operations
Figure 4-1: Correct filler/centrifuge alignment