Loading and Removing the Disposable Tubing Set and Other Common Operations
The Spectra version 7.0 screen:
Label LRS Turbo product <1E6 WBC
content.
Label the product bags to notify the laboratory that this product is
leukoreduced.
– or –
If the platelet yields could be less than predicted and/or events
occurred during the procedure that may cause WBC
contamination, one of the following screens appears:
The Spectra version 5.1 screen:
Measure platelet and WBC content of
this product!
The Spectra version 7.0 screen:
Measure platelet and WBC content!
(CODE)
Label the product to notify the laboratory to measure this product
for platelet and WBC content.
The second screen above appears in LRST procedures. See
"Measuring Platelet and WBC Content Codes" in Chapter 5 of the
Platelet Collection Guide for the codes and their meanings.
WBC content is usually below 1 x 10
should be verified to ensure the product meets local standards for
leukoreduction.
4-14
Press CLEAR
Press CLEAR
Press CLEAR
6
, but actual WBC content
6
The following screen appears:
Clamp and disconnect collection bags
ELP, LRS, and LRST procedures: IMPORTANT: Seal the collect
line above the connection and remove the collection bags. For
instructions on collecting a multiple platelet product, see "Preparing
a Multiple Platelet Product" in Chapter 5 of the Platelet Collection
Guide.
If you concurrently collected plasma, seal the plasma line and
remove the plasma product bag.
Platelet Depletion procedures: IMPORTANT: Close the slide
clamps on the line to the platelet bags.
MNC, PMN, and AutoPBSC procedures: IMPORTANT: Seal the
collect line and remove the collection and plasma bags.
WBC Depletion and LPE procedures: IMPORTANT: Clamp or
seal the collect line and remove the collection bag(s).
TPE procedures: IMPORTANT: Close the clamp on the line
leading to the plasma bag and replacement fluid bags.
Dual-Needle procedures: Disconnect the access needle and place it
in an appropriate needle disposal container, or disconnect the access
line from the vascular access device.
7
Press CLEAR.
During Rinseback, the Spectra system displays several operation screens,
and pumps saline to the channel to flush RBCs to the donor/patient.

Disconnecting the Donor or Patient

Follow the instructions in the procedural guides.
®
COBE
Spectra™ Apheresis System • Essentials Guide