Preface
For the Seal Safe System
Warning: To avoid the possibility of receiving a radio frequency
(RF) burn at the needle entry point, DO NOT activate the Seal Safe
system sealing head within 3 inches (8 cm) of the access or return
needle.
Warning: To avoid possible electric shock, DO NOT submerge
the Seal Safe system cutter/sealer head in any kind of liquid.
Warning: Separate the Seal Safe system cutter/sealer head
from the radio frequency (RF) interconnect cable before cleaning.
Failure to do so could result in a serious RF burn during cleaning.
Warning: To avoid the possibility of receiving a radio frequency
(RF) burn, DO NOT place fingers within 1 inch (2.5 cm) of the
Seal Safe system sealing jaws while sealing.
Warning: To avoid the possibility of a severe burn, DO NOT
touch the metal section on the lever of the Seal Safe system.
Grasp the lever on the plastic-coated section only.
Warning: DO NOT release the Seal Safe system lever or try to
separate the tubing until 1 second after the indicator light turns
off. Releasing the lever sooner could result in an inadequate seal,
subjecting the blood product to non-sterile conditions.
Warning: To avoid electric shock, DO NOT open the head of the
Seal Safe system during servicing of the cutter/sealer head. Refer
servicing to a qualified technician.
1-12
Warning: Since the Seal Safe system complies with the most
restrictive electromagnetic emissions limits imposed on electrically
powered equipment, pacemaker users need not exercise unusual
precautions when in the vicinity of the Seal Safe system. However,
pacemaker users should review their pacemaker manufacturer's
instructions for use for any cautions or contraindications with
respect to use of the pacemaker in the vicinity of equipment such
as the Seal Safe system, which may emit electromagnetic energy.

Cautions

Below are cautions that appear in the manual. Many of the cautions
instruct you to contact Gambro BCT. U.S. customers should call
Gambro BCT Customer Support. Customers outside the U.S. should
contact their local Gambro BCT representative.
For the Spectra System
Caution: Rx only.
Caution: Before installing a filler or loading a disposable tubing
set, inspect the filler and visible interior of the centrifuge chamber
for fluid spills or structural damage. Ensure the metal band around
the perimeter of the filler is centered and secure.
Caution: Disposable tubing sets are for single-use only. DO NOT
resterilize.
®
COBE
Spectra™ Apheresis System • Essentials Guide