CENTRIFUGE COVER OPEN!

CENTRIFUGE COVER OPEN!
Close cover.
Alarm LED is red. Pumps stop. Centrifuge stops.
Probable Cause
Cover left open when centrifuge should be spinning
Cover safety latch not engaged, even though the
door appears to be closed
Dust buildup on the front door sensor
Door switch or wiring malfunction
®
COBE
Spectra™ Apheresis System • Essentials Guide
CONTINUE
Close the centrifuge cover. To continue the procedure, press CONTINUE.
Wiggle the cover. Press UNLOCK COVER and open the cover. Close the cover and door again.
Remove the dust buildup from the front door sensor. For instructions, see "Every Month" on page 10-3.
Note the alarm message. Discontinue the procedure.
Press CHANGE MODE, then press 4 to select "Rinseback." To override the alarm during Rinseback mode, press
CONTINUE.
Notify a qualified service technician of the alarm message.
To override the alarm, press CONTINUE .
.
Action
General Alarms and Troubleshooting
12-21