General Alarms and Troubleshooting

FAILURE #3: __Display Comm__ __ _

FAILURE #3: __Display Comm ___ ____ __
Record #.CONTINUE to retry or Rinseback.
Alarm LED is red. Pumps stop. Depending on alarm, the centrifuge
stops and/or the return line valve closes.
Probable Cause
Brief power disruption
Internal system malfunction
12-32
Turn off the system, wait 5 seconds, then turn it on again. The "POWER INTERRUPTED!" alarm appears. To continue the
procedure, press CONTINUE.
Record the entire alarm message. The code and text data it contains could help a qualified service technician diagnose the
cause of the problem.
Press CHANGE MODE, then press 4 to select "Rinseback." To override the alarm during Rinseback mode, press
CONTINUE.
If you cannot override the alarm during Rinseback mode, turn off the system. See the "Manual Rinseback Procedure" on
page 11-8.
Notify a qualified service technician of the alarm message.