Check the CCM for proper loading

Check the CCM for proper loading
Press CONTINUE
Alarm LED is red.
Probable Cause
Cuvette not properly installed in the CCM
Incorrect disposable tubing set selected during data entry

COLLECT PUMP ERROR!

COLLECT PUMP ERROR! Check for proper
loading or leaks.
Alarm LED is red. Pumps stop.
Probable Cause
Tubing incorrectly positioned in the collect pump or return
pressure sensor
Plasma, return, waste, RBC, or collect valve incorrectly
loaded
Incorrect disposable tubing set selected during data entry
®
COBE
Spectra™ Apheresis System • Essentials Guide
Ensure the cuvette is properly installed in the CCM. Press CONTINUE.
Press CHANGE MODE, press 1 to select "Load Set," choose the correct disposable tubing set, and reprime.
CONTINUE
Ensure the collect pump, valves, and return pressure sensor are loaded correctly.
Press CONTINUE.
Check for proper valve loading.
Press CHANGE MODE, press 1 to select "Load Set," choose the correct disposable tubing set, then reprime.
Press CONTINUE to remove the alarm message.
.
Action
To retry the collect pump test, press CONTINUE.
This alarm occurs only during Prime mode when the collect pump fails
to develop sufficient pressure at the return pressure sensor.
Action
Note: For instructions on how to reposition the diaphragm and reload the return pressure sensor,
see "Repositioning the Return Pressure Sensor Diaphragm" on page 11-10.
General Alarms and Troubleshooting
12-25