The Spectra system uses four different fillers (see Figure 2-6). The system
performs each apheresis procedure using a specific filler. For information
about which filler to install, see "Removing and Installing the Filler" on
page 4-2, or refer to the procedural guides.
Figure 2-6: Fillers from left to right: single-stage, dual-stage, dual-stage with LRS bracket, and dual-stage LRS Turbo
Return Flow Controller
Caution: You must use the Return Flow Controller to perform a
Single-Needle procedure. DO NOT attempt to perform a Single-
Needle procedure without a Return Flow Controller and
appropriate disposable tubing set.
Use the COBE Spectra Return Flow Controller to perform Single-
Needle ELP, LRS, LRST collections and Single-Needle TPE procedures.
The Return Flow Controller has two functions:
Holds the Single-Needle bag, which contains blood components the
system removed from the donor/patient during the Draw phase of a
Single-Needle procedure. The system returns the components to the
donor/patient during the Return phase.
®
COBE
Spectra™ Apheresis System • Essentials Guide
Provides controlled pressure on the Single-Needle bag during the
Return phase of a Single-Needle procedure.
For an explanation of the symbols that appear on the Return Flow
Controller's front cover, see Figure 2-12 on page 2-20. This information
also appears on the permanent label on the back cover of the unit.