General Alarms and Troubleshooting

Service mode enabled.

Service mode enabled.
Stop procedure. Notify Field Service.
Warning LED is yellow.
Probable Cause
Service mode enabled or system ERROR alarms disabled
Total plasma collected (collect and plasma bags) exceeds specified limit.
Total plasma collected (collect and
plasma bags) exceeds specified limit.
Warning LED is yellow.
Probable Cause
Operator increased the plasma collection volume beyond
the value configured for total plasma collection.
12-62
.
Note the warning message. Wait 60 seconds for the yellow Warning LED to stop flashing. Press CLEAR, then
press ENTER.
If the warning reappears, discontinue the procedure. Notify a qualified service technician of the alarm
message.
If the warning does not reappear, the Service mode is disabled for this procedure. Continue the procedure.
To override the warning, press CLEAR. Lower the plasma or platelet collect volume, or shorten the procedure
time.
If you override the warning and make no adjustments, the system will collect the volume you entered.
Caution: While in Service mode, DO NOT use the Spectra
system for any procedure.
Action
Note: Check regulations for how much plasma you can remove
from a donor.
Action
®