Place the lines in the collect/replace and plasma valves (see Figure
4-6).
Figure 4-6: Placing lines in collect/replace and plasma valves
Tip: Use a flossing action when possible to press tubing
into valves, sensors, and fittings. Ensure the lines are
inserted as far back into the fitting as possible.
17
TPE and RBCX procedures: Lay the replacement fluid spike
assemblies over the top of the system.
18
Place the return pressure sensor in its housing. Push downward and
turn clockwise to lock in place.
®
COBE
Spectra™ Apheresis System • Essentials Guide
Loading and Removing the Disposable Tubing Set and Other Common Operations
19
TPE, Platelet Depletion, AutoPBSC, ELP, LRS, and LRST
procedures: Install the collect concentration monitor (CCM)
cuvette as follows:
Note: DO NOT touch the cuvette. Fingerprints can cause
inaccurate CCM readings and inconsistent spillover
detection.
Pull the housing out and turn it counterclockwise to lock it
a
into the load (open) position.
b
While holding the tubing above and below the cuvette, slide
the cuvette into position in the CCM (see Figure 4-7). Ensure a
flat side of the cuvette is parallel to the front panel.
Figure 4-7: Sliding cuvette into CCM
c
Turn the housing clockwise and gently lower it over the cuvette
to lock it into position.
Note: AutoPBSC procedures The cuvette slides freely
on the collect line.