Ratio configuration was changed last run.

Ratio configuration was changed
last run. Please review.
Warning LED is yellow.
Probable Cause
The configured inlet:AC ratio was changed for the last
procedure

RBCs DETECTED.

RBCs DETECTED.
Check RBC valve for proper loading.
Warning LED is yellow.
Probable Cause
Tubing incorrectly positioned in the RBC line valve
Dirty RBC detector
Reprime selected after blood entered the disposable tubing set
Internal system malfunction
®
COBE
Spectra™ Apheresis System • Essentials Guide
CLEAR
1
To remove the warning message and display the inlet:AC Ratio configuration screen, press CLEAR.
2
To continue to use the configured inlet:AC ratio, press ENTER.
– or –
Change the configured inlet:AC ratio, then press ENTER.
Ensure the tubing is placed correctly in the RBC line valve (pressed firmly in the bottom).
Clean the sensor. See "Cleaning the sensors" on page 10-2.
Stop the procedure and disconnect the donor/patient.
Procedure can continue. The system uses time-out backup to close the waste divert valve and open the
return line valve.
Verify this action. If it does not take place, manually close the waste divert valve and open the return line
valve.
.
Action
To override the warning for 60 seconds, press CLEAR.
This warning occurs only during Prime mode when RBCs have been
detected at the wrong time.
Action
General Alarms and Troubleshooting
12-51