General Alarms and Troubleshooting

Plasma and collect pumps running faster than inlet.

Plasma and collect pumps running
faster than inlet.
Warning LED is yellow.
Probable Cause
The combined plasma pump and collect pump flow rates
remove blood from the channel faster than it is being
pumped in

Plasma collection volume exceeds pump speed limit.

Plasma collection volume exceeds pump
speed limit. Plasma volume reduced.
Warning LED is yellow.
Probable Cause
Enabling the High-Flow option to improve collection yield
– or –
Operator entered a large plasma target value late in Run
mode
12-46
This warning occurs only during Manual mode.
To override the warning, press CLEAR. Reduce the plasma pump and collect pump flow rates.
Target plasma volume reduced to the maximum the system can collect in
the remaining procedure time.
This warning occurs only during concurrent collection of plasma.
To remove the warning message, press CLEAR. Complete the procedure using the target plasma volume the
system calculated.
Action
Action
®