How To Remove The Cover To Connect The Battery - Gambro AK 95 S Service Manual

Dialysis machines
Table of Contents

Advertisement

How to remove the cover to connect
the battery.
1.
Loosen the three cover bolts, accessible
underneath the fluid part.
2.
Move the cover upwards and then back-
wards.
3.
Pull the cover to the right just 10 mm (if you
are positioned in front of the AK 95 S).
4.
Turn the cover backwards counter-clockwise
to pass the circuit boards and the cover can
be removed.
5.
Connect the rechargeable battery to board
connector P205 on the Protective Driver
board.
AK 95 / AK 95 S Service Manual
HCEN 9133 Rev. 01. 2004
A
Cover
Front
A
Acet ate
A
Installation Guide
B
A
1:5