Structure of the command sequences for data communication

The interface transmission speed can be selected within the defined baud rates (9,600, 19,200,
38,400, 57,600 Baud) at the touchscreen of the S
Setup"
Structure of the command sequences for data
communication
All characters sent and received in the data exchange between a PC and the incubator
S
TERI-CYCLE
terminal.
This ensures that communication can be set up, controlled, and programmed easily.

Description of protocol

Character coding
ASCII characters, capital letters are not allowed.
Reading parameters:
Query:
or:
Response:
where:
Description of the response elements:
aaaa
bb
cc
Example of a software version query (50111927):
Query:
Response:
Writing parameters:
Command:
Response:
where:
Response with an error message:
Response:
i160 / i250
S
12-4
TERI-CYCLE
on
page
6-17).
i160 / S
i250 are ASCII characters that can be displayed on a conventional
TERI-CYCLE
?:aaaa:bb::cc<CR>
?:aaaa:bb:XXXX:cc<CR>
!:aaaa:bb:XXXXX:cc<CR>
aaaa = parameter adress
bb =amount of payload in this telegram (00 – ff )
cc = checksum: Inverted XOR of all bytes without
checksum and <CR>
XXXX = bb-bytes user data
parameter address
amount of payload in this telegram (00 – ff )
Checksum: Inverted XOR of all bytes without
checksum and <CR>
?:0001:00::cc<CR>
!:0001:08:50111927:cc<CR>
!:aaaa:bb:XXXXX:cc<CR>
!:aaaa:bb::cc<CR>
aaaa = parameter adress
bb =amount of payload in this telegram (00 – ff )
cc = checksum: Inverted XOR of all bytes without
checksum and <CR>
XXXX = bb-bytes user data
!:aaaa:bb:XX:cc<CR>