Auto-Start Function - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

6
Handling and control

auto-start function

Note Factory settings:
Depending on the type of the O
factory:
Control range I: 1 % - 21 %
Control range II: 5 % - 90 %
Usage of the process gases:
For set values below 21 % O
For set values above 21 % O
For set values of 21 % O
The control loop error monitoring system (optional) is enabled.
Note
Please make sure that the sensor has been correctly plugged into the socket. Improper installation of
the sensor may cause corrosion at the contacts and calibration errors during the auto-start routine.
You can verify the proper function by simply activating the sensor. When no error message occurs
within 10 minutes, the unit is ready for the auto-start routine.
auto-start function
The auto-start function is a fully automated routine for startup and subsequent adjustment of the CO
sensing circuits. After the start, the device control adjusts the temperature to the set value while
humidity is generated. When temperature and relative humidity have reached constant values, the
CO
measuring system is automatically adjusted to these values, and the workspace is supplied with
2
the preset quantity of CO
i160 / i250
S
6-12
TERI-CYCLE
sensor, one of the two O
2
, the device must be connected to a nitrogen supply system.
2
, the device must be connected to an oxygen supply system.
2
, the O2 control is switched off and error monitoring is disabled as well.
2
.
2
control ranges has been preset at the
2
Thermo Scientific
2