Overview Of General Parameters (Addresses 0Xxx); Reading General Parameters; Overview Of Incubator Parameters (Addresses 2Xxx) - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

Description of the response elements:
aaaa
bb
cc
Example of an unknown command:
Query:
Response
Significance of the two bytes in the error message:
Failure message
?0
?1
?2
?3

Overview of general parameters (addresses 0xxx)

General parameters are system values such as date, time, and motherboard version number.

Reading general parameters

Address
0001
0010
0011
0012

Overview of incubator parameters (addresses 2xxx)

Incubator parameters are divided into:
• the basic parameters of the three control loops temperature, CO
• the internal function parameters of operating functions and data logging.
Thermo Scientific
parameter address,
amount of payload (always 02)
Checksum: Inverted XOR of all bytes without
checksum and <CR>
XX = 2 bytes error message (see table below)
?:0005:00::cc<CR>
!:0005:02:?1:cc<CR>
Description
Error in telegram structure or check sum
Unknown error or unknown parameter
Internal memory error
Data error (value not within set limits)
Description
Mainboard version number
Date and time display
[hours:minutes:seconds];[day:mon
th:year]
Date
[day:month:year]
Time
[hours:minutes:seconds]
12
Annex: Data communication
Overview of general parameters (addresses 0xxx)
Remarks
8 places
17 bytes / decimal value
with the format xx:xx:xx;xx:xx:xx
8 bytes / decimal value
with the format xx:xx:xx
8 bytes / decimal value
with the format xx:xx:xx
and O
,
2
2
S
TERI-CYCLE
i160 / i250
12-5