Co2 Supply; O2 Supply; Figure 3-5. Air Duct - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

Figure 3-5. Air duct

The air duct consists of two shaped pieces of sheet metal:
1. Ceiling duct
2. Rear wall duct
3. Back wall of the workspace
The air duct and HEPA-filter can be installed and removed without any tooling.
CO
supply:
2
To ensure the growth conditions for the cell and tissue cultures, the workspace is supplied with CO
The pH of the bicarbonate-buffered culture media largely depends on the CO
workspace atmosphere.
The CO
The supplied CO
• Purity 99.5 % min
• medical gas quality.
O
supply:
2
If the CO
oxygen (optional).
The O
2
Operation with higher oxygen concentrations requires that the fire protection caution notes in
in the section
Thermo Scientific
1
(1/Figure
3-5)
(2/Figure
3-5)
(3/Figure
content of the workspace atmosphere can be controlled within a range of 0-20 %.
2
must have one of the following quality characteristics:
2
incubator is to be operated with more than 21 % oxygen, the workspace is supplied with
2
content of the workspace atmosphere can be controlled within a range of 21 % to 90 %.
Safety notes on gases
2
3-5)
on page
7
be heeded.