Replacing The Gas Inlet Filter; Figure 9-7. Installing The Hepa-Filter - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

6. Turn the airbox upside down, unlatch the tabs
filter tabs
7. Unlatch the tabs on the right-hand side
corresponding slots provided in the HEPA-Filter
1
2

Figure 9-7. Installing the HEPA-filter

8. Install the new HEPA-filter in airbox
9. Place the airbox on its seat in the water reservoir cover panel.
10. If the unit has been operated without the HEPA-filter previously, activate the HEPA-filter in the
user configuration according to instructions
page
11. If necessary, set the reminder interval for the HEPA-filter replacement in the user configuration
according to instructions
HEPA-filter replacement may be set between 1 and 12 months. We recommend setting this
parameter to the factory preset value of 6 months.

Replacing the gas inlet filter

The gas inlet filter (CO
thread socket at the control box.
Thermo Scientific
(6/Figure
9-7).
7
8
6-39.
"Reminder interval setting"
-/O
-/N
2
2
2
(5/Figure
(3/Figure
9-7) of the airbox
(4/Figure
3
4
5
6
(1/Figure
9-7) and engage the locking tabs.
"Activating / de-activating the HEPA-filter:"
on
supply) has a plastic thread and is screwed by hand into the female
Replacing the gas inlet filter
9-7)on the left-hand side from the HEPA
(1/Figure
9-7) from the
9-7).
4
CLICK
page
6-26. The reminder interval for
i160 / i250
S
TERI-CYCLE
9
Maintenance
3
on
9-9