Figure 8-3. Air Duct - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

8
Cleaning and disinfection
Wipe/spray disinfection
A
E

Figure 8-3. Air duct

5. Unhinge the rear part of the air duct
6. Remove pre-filter
The pre-filter is autoclavable.
7. Remove the cover panel from the water reservoir.
Cleaning the workspace and accessories:
1. Thoroughly remove dirt residues and deposits using a solution of tepid water and dishwashing
agent.
2. Wipe the surfaces clean using a clean cloth and plenty of clear water.
3. Remove the cleaning liquid from the water reservoir and wipe all surfaces of the workspace
thoroughly dry.
4. Wipe accessories thoroughly dry.
Final disinfection:
1. Spray the surfaces of the workspace, the shelving system and parts removed in previous steps with
disinfectant one more time and wipe dry.
2. Allow the disinfectant to react as specified by the manufacturer.
3. Reinstall the shelf system and the removed components.