8
Cleaning and disinfection

Decontamination procedures

Decontamination procedures
The operator must prepare hygiene regulations for the decontamination of the device in accordance
with the application of the device.
The following disinfection procedures are compatible with the device:
Wipe/spray disinfection:
The wipe/spray disinfection is used as the standardized manual disinfection procedure for the device
and for all accessories.
Steri-run decontamination routine:
The steri-run decontamination routine uses an automated program cycle to decontaminate the
complete workspace including the shelf system and the sensors.

Preparing for disinfection or steri-run

1. Remove all samples from the workspace and store them in a safe place.
2. Procure a collecting vessel of sufficient capacity.
3. Place open end of hose into collecting vessel and lock drain hose valve safely into receptacle
4. The water reservoir starts draining.

Figure 8-1. Drain valve of water reservoir

5. Allow the water reservoir to drain completely into the collecting vessel.
6. Wipe up residual water with a cloth.
7. Pull airbox
8. Take HEPA-filter
i160 / i250
S
8-2
TERI-CYCLE
(1/Figure
8-2) out of its seat on the water reservoir cover panel and remove it.
(2/Figure
8-2) out of airbox
(1/Figure
8-2) and reinstall the airbox.
Thermo Scientific