O2 Control Loop And Water Level; Data Logger Structure - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

Bit
0x0004
0x0010
0x0020
0x0040
0x0080
0x0200
0x0400
O
control loop and water level:
2
Bit
0x0001
0x0002
0x0004
0x0020
0x0040
0x0080
0x0200
0x0400
Bit
0x0001

Data logger structure

The data logger can store up to 10,000 events. Depending on the setting for the logging cycle (in
sections of seconds), e.g. for a value of 10,000 s (default value), the events of about 5 days can be
stored.
The data logger stores the following information:
• important user actions, system events, and error messages,
• measured data of the three control loops during incubation operation.
The data logger can be queried using the following commands:
Query:
Query:
Query:
Thermo Scientific
CO
control loop error status
2
Actual value low
Calibration values too high/low
Error communication (to sensor)
Error communication (to gas cylinder changing unit)
No gas available, bottle A and B empty
Cylinder A empty
Cylinder B empty
O
control loop error status
2
Sensor breakage
Actual value high
Actual value low
Error communication (to sensor)
Gas cylinder changeover switch does not communicate
No gas available, bottle A and B empty
Cylinder A empty
Cylinder B empty
Water level error status
No water
?:2400:00::cc<CR>
Set the data logger pointer to the oldest entry and output the
first data sets.
?:2401:00::cc<CR>
Output of the following data sets, the reading pointer automatically moves
gradually from the older entries to the current entries.
?:2402:00::cc<CR>
12
Annex: Data communication
Data logger structure
i160 / i250
S
TERI-CYCLE
12-11