11
Specifications
STERI-CYCLE i160
S
i160
TERI-CYCLE
Name
CO
gas supply system
2
Gas purity
Primary pressure
Measurement and control range
Control deviation over time
CO
measuring cell
2
Accuracy at 37 °C and 5% CO
O
gas supply system
2
Gas purity
Primary pressure
Measurement and control range
Control deviation over time
O
measuring cell
2
Accuracy at 37 °C and 21% O
Electrical
Rated voltage
Rated frequency
Degree of protection (IEC 60529)
Protection class
Overvoltage category (EN 61010)
Pollution degree (EN 61010)
Rated current
Circuit breaker
i160 / i250
S
11-4
TERI-CYCLE
Unit
%
bar
Vol -%
Vol -%
%CO
2
%
bar
Vol -%
Vol -%
%O
2
V
Hz
A
Value
min. 99.5 or med. quality
min. 0.8 - max. 1 (11.6-14.5 psi)
0...20
± 0.1
± 0.3
2
min. 99.5 or med. quality
min. 0.8 - max. 1 (11.6-14.5 psi)
1... 21 or 5...90
± 0.2
± 0.5 (option: 1.....21% O
2
± 2,0 (option: 5.....90% O
1/N/PE 230 V, AC (± 10%)
1/N/PE 220 V, AC (± 10%)
1/N/PE 120 V, AC (± 10%)
1/N/PE 100 V, AC (± 10%)
50/60
IP 20
I
II
2
230 V:
Decontamination: 4.6
Incubation: 2.4
220 V:
Decontamination: 4.4
Incubation: 2.3
120 V:
Decontamination: 8.3;
Incubation: 4.6
100 V:
Decontamination: 7.2;
Incubation: 3.9
16 A
)
2
)
2