Handling and control
User configuration

Figure 6-38. Low humidity setting

1. To toggle between two states:
• Press the + key.
or
• Press the - key.
2. To accept and save the change:
• Press the Enter key.
The system returns to the Options menu.
Upon return to the main menu the Low humidity icon appears.
Note Low humidity:
The enabling/disabling of the Low humidity function is entered into the event list.
Gas tight screen setting
As the aperture cross-sections are smaller when accessing the samples, devices equipped with the
optional gas-tight screen achieve shorter recovery times for the incubation parameters:
• Workspace temperature,
• CO
• O
2
• relative humidity.
After conversion of the device, the device control system must be adjusted to the gas-tight screen
option.
Note Malfunction:
The conversion to gas-tight screen causes a change of the control parameters.
Failure to set the gas-tight screen function in accordance with the actually installed door version
may impair the incubation performance.
Gas tight screen setting
1. Press the Menu key.
2. Select the menu command Options.
i160 / i250